EnglishPortugueseSpanish

Request a Quote / Contact Us!

    Manufacturer

    Part Number or Numbers

    GTEM CELL 1800 K FROM GTEM ITALIA

    GTEM 1800 K GTEM Cell from GTEM Italia

    For emission and immunity testing according to IEC/EN 61000-4-20
    GTEM Italia GTEM 1800 K

    The GTEM-1800 is a TEM waveguide with the upper frequency limit extended to the GHz range. It is under consideration as an alternative measurement facility for both radiated emission and immunity measurements. It is included in the standard IEC 61000-4-20 “Emission and Immunity Testing in Transverse Electromagnetic (TEM) Waveguides”.


     

    GTEM ITALIA GTEM 1800 K EUT Size and Placement Image

     

    The GTEM 1800 K ships in a Kit and requires assembly on-site.


    Main Technical Specifications / Max EUT Size

    Frequency range

    For emissions:

    9KHz – 3 GHz

    For immunity applications:

    80MHz – 6GHz

    * DC-20GHz optional

    Septum Height [mm]:

    1800 mm

    Max EUT Size (LxWxH)cm

    132x132x100cm

    Defined test Vol. within 6dB

    58x58x58cm

    Typical VSWR

    1:1.2

    (for immunity test range)

    Typical VSWR at critical frequency

    <1:1.6

    (for immunity test range)

    Max Input power, W continuous/*pulsed

    1,5 kW/2.5 kW

    Input connector

    7/16“ or N UG-21

    Nominal impedance

    50 Ohm

    Absorbers

    55cm/21,6” Truncated pyramid

     

    Mechanical Dimensions

    Outer LxWxH [cm]:

    820 x 428 x 283cm + 25cm Trolley

    Door WxH [cm]:

    Standard: 100 x 160cm * Optional: 140x140cm

    Window in the door (WxH)

    * Optional: circular Diam.200mm.

    Weight Kg. Approx

    Approx. 1600kg

    Wheeled undercarriage

    No. 16

     

    Electrical Specifications

    Mains connectors

    Fix/CEE

    Main Switch

    magneto-thermal 16A mono phase

    Input Socket plug

    16 AC IEC Type ( mono phase + ground)

    Output Socket EUT tape

    16 AC SHUCO type (mono phase + ground)

    Additional EUT sockets

    Optional

    Ground connection

    M6 bolt,(it is required to be always connected)

    EMI AC Line filter (mono phase + ground)

    16A 250V n.2 terminals +ground

    EMI AC/DC secondary filter

    10A 250V AC/DC N.2 terminals

    Channel for fiber optic leads

    3 couples (18GHz Cut-off frequency)

    RF feed-thru connector

    N.1 N-N female

    RF feed-thru SMA type connectors

    N.2 SMA-SMA female

    Technical panel pre-drilled for options

    supplied

     

    Options

     

    Midia filter unit

    *Optional: N.1 RJ45, N.1 DB9-RS232, N.2 USB-2, N.1
    DC filter.

    Video camera system

    * Optional

    Indoor lighting

    10W LED

    Manipulator X,Y,Z

    Motorized USB, Manual, Remote controlled

     

    Frankonia Group Open TEM-Cells GTEM Italia GTEM 1800 K Data Sheet

    Frankonia Group Open TEM-Cells GTEM Italia GTEM 1250 to GTEM 2500 Specifications Matrix

    Frankonia Group Open TEM-Cells GTEM Italia GTEM Calculating Power Required for GTEM250 400 500 750 1000 1250 1500 1750 2000 Data Sheet

    Frankonia Group Open TEM-Cells GTEM Italia GTEM Options and Accessories Matrix

    Frankonia Group Open TEM-Cells GTEM Italia GTEM Superior Technology over the Competition