EnglishPortugueseSpanish

Request a Quote / Contact Us!

Manufacturer

Part Number or Numbers

HILO – Coupling Network (CDN)

Coupling network (CDN)

• Hilo-Test Coupling Network (CDN)

• CDN 304 COUPLING / DECOUPLING NETWORK

• CDN 504 sym (Symmetrical) Coupling – /Decoupling Network

• CDN 508 sym (Symmetrical) Coupling – /Decoupling Network

• CDN 2012 Coupling Clamp

• CDN 2402 Capacitive Coupling-/Decoupling Network

• CDN 2410 Capacitive Coupling-/Decoupling Network

• CDN 2802 Capacitive Coupling-/Decoupling Network

• CDN 4416 Capacitive Coupling-/Decoupling Networks

• CDN 4416 B Capacitive Coupling-/Decoupling Network

• CDN 4432 Capacitive Coupling-/Decoupling Network

• CDN 4463 Capacitive Coupling-/Decoupling Network

• CDN 5404 Capacitive Coupling-/Decoupling Network

• CDN 5416 Capacitive Coupling-/Decoupling Network

• CDN 5416 B Capacitive Coupling-/Decoupling Network

• CDN 5432 Capacitive Coupling-/Decoupling Network

• CDN 5463 Capacitive Coupling-/Decoupling Network

• CDN 6416 Capacitive Coupling-/Decoupling Network

• CDN 6432 Capacitive Coupling-/Decoupling Network

• CDN 6463 Capacitive Coupling-/Decoupling Network

• CDN 7416 Capacitive Coupling-/Decoupling Network

• CDN 7432 Capacitive Coupling-/Decoupling Network

• CDN 7463 Capacitive Coupling-/Decoupling Network

• CDN 10216 Capacitive Coupling-/Decoupling Network

• CDN 10416 Capacitive Coupling-/Decoupling Network

• CDN 10432 Capacitive Coupling-/Decoupling Network

• CDN 10463 Capacitive Coupling-/Decoupling Network

• CDN 12216 Capacitive Coupling-/Decoupling Network

• CDN 12416 Capacitive Coupling-/Decoupling Network

• CDN 12432 Capacitive Coupling-/Decoupling Network

• CDN 12463 Capacitive Coupling-/Decoupling Network

• CDN 44100 Capacitive Coupling-/Decoupling Networks

• CDN 44250 Capacitive Coupling-/Decoupling Network

• CDN 54100 Capacitive Coupling-/Decoupling Network

• CDN 54200/ CDN 74200/CDN 104200/ CDN 124200 Capacitive Coupling-/Decoupling Networks

• CDN 54250 Capacitive Coupling-/Decoupling Network

• CDN 64100 Capacitive Coupling-/Decoupling Network

• CDN 64250 Capacitive Coupling-/Decoupling Network

• CDN 74100 Capacitive Coupling-/Decoupling Network

• CDN 74250 Capacitive Coupling-/Decoupling Network

• CDN 104100 Capacitive Coupling-/Decoupling Network

• CDN 104250 Capacitive Coupling-/Decoupling Network

• CDN 124100 Capacitive Coupling-/Decoupling Networks

• CDN 124250 Capacitive Coupling-/Decoupling Network

• EFTC 2012 Capacitive Coupling Clamp

 

 

Hilo-Test Coupling Network (CDN) 

HILO-TEST offers a large number of coupling / decoupling networks (CDN) for single-phase and 3-phase power supply lines and signal data lines and telecommunication lines. In addition to offering numerous versions for current EMC Standards (IEC 61000-4-4 and IEC 61000-4-5), HILO-TEST also supplies special variants for higher voltages (e.g. 3×690 VAC or up to 1 KV DC) and high currents (up to 200 A or more), as well as for surge voltages up to 12.kV.

The related standards are formally valid IEC 61000-4-4 and IEC 61000-4- standards.

 

CDN 304 COUPLING / DECOUPLING NETWORK 

Summary:

• 100 V= / 1 A

• max. test voltage: 5 kV, 1.2/50μs,10/700 μs

• max. test current: 2 KA, 8/20 μs

 

Designed for applying the impulse output voltage of the Combination Wave Generator to interconnection lines of the equipment under test, acc. to IEC 1000-4-5.

 

Related Standards:

IEC: IEC 61000-4-05

 

 CDN 304 Capacitive Coupling-/Decoupling Network Data Sheet
 

 

CDN 504 sym (Symmetrical) Coupling-/Decoupling Network 

CDN 504 sym (Symmetrical) Coupling-/Decoupling Network

Summary:

Rated voltage 80V=
Surge: 5.0 kV, 1.2/50 μs
Surge: 5.0 kV, 10/700 μs
Coupling Impedance, selectable

 

Designed for applying the impulse output voltage of the Combination Wave Generator to interconnection lines of the equipment under test, acc. to IEC 61000-4-5, Fig10

 

Related Standards: IEC: IEC 61000-4-05

 

CDN voltage: up to 5kV

CDN usage: Surge (CWG), Data- / signal lines

 

 CDN 504 sym (Symmetrical) Coupling – /Decoupling Network Data Sheet

 

CDN 508 sym (Symmetrical) Coupling-/Decoupling Network 

CDN 504 sym (Symmetrical) Coupling-/Decoupling Network

Summary:

Rated voltage 80V=
Surge: 5.0 kV, 1.2/50 μs
Surge: 5.0 kV, 10/700 μs
Coupling Impedance, selectable

 

Designed for applying the impulse output voltage of the Combination Wave Generator to interconnection lines of the equipment under test, acc. to IEC 61000-4-5, Fig10

 

Related Standards: IEC: IEC 61000-4-05

CDN voltage: up to 5kV

CDN usage: Surge (CWG), Data- / signal lines

 

 CDN 508 sym (Symmetrical) Coupling – /Decoupling Network Data Sheet

 

CDN 2012 Coupling Clamp 

HILO-TEST CDN 2012 Coupling Clamp

 

Summary:

• Coupling clamp for automotive electronics

• Pulse injection on signal and control lines

 

The coupling clamp CDN 2012 conforms exactly to the requirements of ISO 7637-3 and other standards, guaranteeing that tests are carried out in strict compliance with the standard. The CDN allows the fast nanosecond pulse bursts ( ISO 3a and 3b) to be injected in cable runs. Very high impedance signal lines are also to be found in vehicles on which the disruptive effects of ISO pulses 1 and 2 can also show up through Capacitive coupling.

 

Related Standards:

ISO: ISO 7637-3

 

 CDN 2012 Capacitive Coupling Clamp Data Sheet
 

 

Hilo-Test CDN 2402 

Hilo-Test CDN 2402

 

Details:

• Data- / signal lines
• Surge: 2.5 kV, 1.2/50 μs
• Burst: 2.5 kV , 5/50 ns
• Typ. 48 V=

 

The capacitive Coupling-/Decoupling Networks CDN 2402/2410 are used in combination with the CE-TESTER and allows superimposition of surge and burst test pulses to up to four I/O signal lines. The test set-up is suitable for surge immunity testing of electronic systems and devices according to IEC 61000-4-4, IEC 61000-4-5 and IEEE 587.

 

Standards:

IEC 61000-4-04, IEC 61000-4-05

 

Associated Downloads

 Hilo-Test CDN 2402 Data Sheet

 

 

Hilo-Test CDN 2410 

Hilo-Test CDN 2410

 

Details:

• Data- / signal lines
• Surge: 2.5 kV, 1.2/50 μs
• Burst: 2.5 kV , 5/50 ns
• Typ. 240V≈

 

The capacitive Coupling-/Decoupling Networks CDN 2402/2410 are used in combination with the CE-TESTER and allows superimposition of surge and burst test pulses to up to four I/O signal lines. The test set-up is suitable for surge immunity testing of electronic systems and devices according to IEC 61000-4-4, IEC 61000-4-5 and IEEE 587.

 

Standards:

IEC 61000-4-04, IEC 61000-4-05

 

Associated Downloads

 Hilo-Test CDN 2410 Data Sheet

 

 

Hilo-Test CDN 2802 

Hilo-Test CDN 2802

 

Details:

• Data- / signal lines • Surge: 2 kV, 1.2/50μs • Burst: 2 kV , 5/50 ns • 8 data lines

 

The capacitive Coupling-/Decoupling Network CDN 2802 is used in combination with the CE-TESTER and allows superimposition of surge and burst test pulses to up to eight I/O signal lines.

 

Standards:

IEC 61000-4-04, IEC 61000-4-05

 

Associated Downloads

 Hilo-Test CDN 2802 Data Sheet

 

 

Hilo-Test CDN 4416 

Hilo-Test CDN 4416

 

Details:

• Power supply lines • 3* 400 V / 16 A • Surge: 5.0 kV, 1.2/50 μs • Surge: 2.5 kA, 8/20 μs • Burst: 5.0 kV, 5/50 ns

 

The capacitive Coupling-/Decoupling Networks CDN 4416/6416/10416/12416 are used in combination with the CE-Tester or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

 

Standards:

IEC 61000-4-04, IEC 61000-4-05

 

Associated Downloads

 Hilo-Test CDN 4416 Data Sheet

 

CDN 4416 B Capacitive Coupling-/Decoupling Network 

CDN 4416 B capacitive Coupling-/Decoupling Network

 

Summary:

• Power supply lines

• 3 * 400V / 25 A

• Burst: 5 kV, 5/50 ns

 

The Capacitive Coupling-/Decoupling Network CDN 4416B is used in combination with a BURST Generator EFTG 4510 or the CE-TESTER and allows superimposition burst test pulses to the 3-phase mains voltage of the device under test.

 

Related Standards:

IEC: IEC 61000-4-04

 

 CDN 4416 B Capacitive Coupling-/Decoupling Network Data Sheet
 

 

Hilo-Test CDN 4432 

HILO-TEST CDN 4432

 

Details:

• Power supply lines • 3* 400 V / 32 A • Surge: 5.0kV, 1.2/50 μs • Surge: 2.5A, 8/20 μs • Burst: 5.0 kV, 5/50 ns

 

The capacitive Coupling-/Decoupling Networks CDN 4432/6432/10432/12432 are used in combination with the CE-Tester or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

 

Standards:

IEC 61000-4-04, IEC 61000-4-05

 

Associated Downloads

 Hilo-Test CDN 4432 Data Sheet

 

CDN 4463 Capacitive Coupling-/Decoupling Network 

CDN 4463 Capacitive Coupling-/Decoupling Network

 

Summary:

• Power supply lines

• 3* 400 V / 63 A

• Surge: 5.0 kV, 1.2/50 μs

• Surge: 2.5 kA, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

 

The Capacitive Coupling-/Decoupling Networks CDN 4463/6463/10463/12463 are used in combination with the CE-Tester or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

 

 CDN 4463 Capacitive Coupling-/Decoupling Network Data Sheet
 

 

HILO-TEST CDN 5404 

Product Summary:

• Ring wave: 4 kV, 100KHz
• 4 data lines (2 Paar)

The capacitive Coupling-/Decoupling Network CDN 5404 is used in combination with the IPG2554 and allows superimposition of damped oscillatory wave up to four I/O signal lines.

For eight data lines, two devices can be used.


Standards:
 IEC: IEC 61000-4-12, IEC 61000-4-18, IEC 60255

 

Associated Downloads

 Hilo-Test CDN 5404 Data Sheet

 

Hilo-Test CDN 5416 

• Power supply lines

• 3* 400 V / 16 A

• Surge: 5.0 kV, 1.2/50 μs

• Surge: 2.5 kA, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

 

The capacitive Coupling-/Decoupling Networks CDN 5416/7416/10416/12416 are used in combination with the Multi-CE5 or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

Standards:

IEC 61000-4-04, IEC 61000-4-05

 

Associated Downloads

 Hilo-Test CDN 5416 Data Sheet

 

Hilo-Test CDN 5416 B

• Power supply lines

• 3* 400 V / 32 A

• Burst: 5.0 kV, 5/50 ns

 

The capacitive Coupling-/Decoupling Network CDN 5416B / 5432B is used in combination with a BURST Generator EFTG-CE5 or the Multi-CE5 and allows superimposition burst test pulses to the 3-phase mains voltage of the device under test.

Standards: IEC 61000-4-04

CDN current: up to 16A

CDN voltage: up to 5kV

CDN usage: Burst (EFTG), Power supply lines

 

Associated Downloads

 Hilo-Test CDN 5416B Data Sheet

 

Hilo-Test CDN 5432 

• Power supply lines

• 3* 400 V / 32 A

• Surge: 5.0kV, 1.2/50 μs

• Surge: 2.5A, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

 

The capacitive Coupling-/Decoupling Networks CDN 5432/7432/10432/12432 are used in combination with the Multi-CE5 or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

Standards:

IEC 61000-4-04, IEC 61000-4-05

 

Associated Downloads

 Hilo-Test CDN 5432 Data Sheet

 

 

Hilo-Test CDN 5463 

• Power supply lines

• 3* 400 V / 63 A

• Surge: 5.0 kV, 1.2/50 μs

• Surge: 2.5 kA, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

 

The capacitive Coupling-/Decoupling Networks CDN 5463/7463/10463/12463 are used in combination with the Multi-CE5 or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

Standards:

IEC 61000-4-4, IEC 61000-4-5 and IEEE 587

 

Associated Downloads

 Hilo-Test CDN 5463 Data Sheet

 

Hilo-Test CDN 6416

HILO-TEST CDN 4416

 

Details:

• Power supply lines • 3* 400 V / 16 A • Surge: 7.0 kV, 1.2/50 μs • Surge: 3.0 kA, 8/20 μs • Burst: 5.0 kV, 5/50 ns

 

The capacitive Coupling-/Decoupling Networks CDN 4416/6416/10416/12416 are used in combination with the CE-Tester or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

 

Standards:

IEC 61000-4-04, IEC 61000-4-05

 

Associated Downloads

 Hilo-Test CDN 6416 Data Sheet

 

 

CDN 6416 Capacitive Coupling-/Decoupling Network 

CDN 6416 capacitive Coupling-/Decoupling Network

 

Summary:

• Power supply lines

• 3* 400 V / 16 A

• Surge: 7.0 kV, 1.2/50 μs

• Surge: 3.0 kA, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

 

The Capacitive Coupling-/Decoupling Networks CDN 4416/6416/10416/12416 are used in combination with the CE-Tester or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

 

Related Standards:

IEC: IEC 61000-4-04, IEC 61000-4-05

 

 CDN 6416 Capacitive Coupling-/Decoupling Network Data Sheet
 

 

CDN 6432 Capacitive Coupling-/Decoupling Network 

CDN 6432 Capacitive Coupling-/Decoupling Network

 

Summary:

• Power supply lines

• 3* 400 V / 32 A

• Surge: 7.0kV, 1.2/50 μs

• Surge: 3.0A, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

 

The Capacitive Coupling-/Decoupling Networks CDN 4432/6432/10432/12432 are used in combination with the CE-Tester or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

 

Related Standards:

IEC: IEC 61000-4-04, IEC 61000-4-05

 

 CDN 6432 Capacitive Coupling-/Decoupling Network Data Sheet
 

 

 

CDN 6463 Capacitive Coupling-/Decoupling Network 

CDN 6463 Capacitive Coupling-/Decoupling Network

 

Summary:

• Power supply lines

• 3* 400 V / 63 A

• Surge: 7.0 kV, 1.2/50 μs

• Surge: 3.5 kA, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

 

The Capacitive Coupling-/Decoupling Networks CDN 4463/6463/10463/12463 are used in combination with the CE-Tester or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

 

 CDN 6463 Capacitive Coupling-/Decoupling Network Data Sheet
 

 

 

CDN 7416 Capacitive Coupling-/Decoupling Network 

CDN 6463 Capacitive Coupling-/Decoupling Network

 

Summary:

• Power supply lines

• 3* 400 V / 16 A

• Surge: 7.0 kV, 1.2/50 μs

• Surge: 3.0 kA, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

 

The capacitive Coupling-/Decoupling Networks CDN 5416/7416/10416/12416 are used in combination with the Multi-CE5 or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

 

Additional Info

Standards: IEC: IEC 61000-4-04, IEC 61000-4-05

CDN current: up to 16A

CDN voltage: up to 7kV

CDN usage: Burst (EFTG), Surge (CWG), Power supply lines

 

 CDN 7416 Capacitive Coupling-/Decoupling Network Data Sheet
 

 

CDN 7432 Capacitive Coupling-/Decoupling Network 

CDN 6463 Capacitive Coupling-/Decoupling Network

 

Summary:

• Power supply lines

• 3* 400 V / 32 A

• Surge: 7.0 kV, 1.2/50 μs

• Surge: 3.0 kA, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

 

The capacitive Coupling-/Decoupling Networks CDN 5432/7432/10432/12432 are used in combination with the Multi-CE5 or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

 

Additional Info

Standards: IEC: IEC 61000-4-04, IEC 61000-4-05

CDN current: up to 32A

CDN voltage: up to 7kV

CDN usage: Burst (EFTG), Surge (CWG), Power supply lines

 

 CDN 7432 Capacitive Coupling-/Decoupling Network Data Sheet
 

CDN 7463 Capacitive Coupling-/Decoupling Network 

CDN 6463 Capacitive Coupling-/Decoupling Network

 

Summary:

• Power supply lines

• 3* 400 V / 63 A

• Surge: 7.0 kV, 1.2/50 μs

• Surge: 3.5 kA, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

 

The capacitive Coupling-/Decoupling Networks CDN 5463/7463/10463/12463 are used in combination with the Multi-CE5 or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

 

Additional Info

Standards:

CDN current: up to 63A

CDN voltage: up to 7kV

CDN usage: Burst (EFTG), Surge (CWG), Power supply lines

 

 CDN 7463 Capacitive Coupling-/Decoupling Network Data Sheet

 

CDN 10216 Capacitive Coupling-/Decoupling Network 

CDN 10216 Capacitive Coupling-/Decoupling Network

Summary:

• Power supply lines

• 240 V / 16 A

• Surge: 10 kV, 1.2/50 μs

• Surge: 5 kA, 8/20 μs

• Burst 5 kV, 5/50 ns

 

The Capacitive Coupling-/Decoupling Network CDN 12216 is used in combination with the Surge generators PG 7-250, PG 10-504 or PG 12-804 and allows superimposition of surge test pulses to the single-phase power supply voltage of the device under test. The test set-up is suitable for immunity testing of electronic systems and devices according to IEC 61000-4-4, IEC 61000-4-5 and IEEE 587.

 

 CDN 10216 Capacitive Coupling-/Decoupling Network Data Sheet

 

 

Hilo-Test CDN 10416

HILO-TEST CDN 4416

 

Details:

• Power supply lines • 3* 400 V / 16 A • Surge: 10.0 kV, 1.2/50 μs • Surge: 5.0 kA, 8/20 μs • Burst: 5.0 kV, 5/50 ns

 

The capacitive Coupling-/Decoupling Networks CDN 4416/6416/10416/12416 are used in combination with the CE-Tester or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

 

Standards:

IEC 61000-4-04, IEC 61000-4-05

 

Associated Downloads

 Hilo-Test CDN 10416 Data Sheet

CDN 10432 Capacitive Coupling-/Decoupling Network 

CDN 10432 Capacitive Coupling-/Decoupling Network

 

Summary:

• Power supply lines

• 3* 400 V / 32 A

• Surge: 10.0kV, 1.2/50 μs

• Surge: 5.0A, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

 

The Capacitive Coupling-/Decoupling Networks CDN 4432/6432/10432/12432 are used in combination with the CE-Tester or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

 

Related Standards:
IEC: IEC 61000-4-04, IEC 61000-4-05

 

 CDN 10432 Capacitive Coupling-/Decoupling Network Data Sheet
 

 

CDN 10463 Capacitive Coupling-/Decoupling Network 

CDN 10463 Capacitive Coupling-/Decoupling Network

 

Summary:

• Power supply lines

• 3* 400 V / 63 A

• Surge: 10.0 kV, 1.2/50 μs

• Surge: 5.0 kA, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

 

The Capacitive Coupling-/Decoupling Networks CDN 4463/6463/10463/12463 are used in combination with the CE-Tester or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

 

 CDN 10463 Capacitive Coupling-/Decoupling Network Data Sheet
 

 

Hilo-Test CDN 12216

HILO-TEST CDN 4416

 

Details:

• Power supply lines • 240 V / 16 A • Surge: 12 kV, 1.2/50 μs • Surge: 6 kA, 8/20 μs • Burst 5 kV, 5/50 ns

 

The capacitive Coupling-/Decoupling Network CDN 12216 is used in combination with the Surge generators PG 7-250, PG 10-504 or PG 12-804 and allows superimposition of surge test pulses to the single-phase power supply voltage of the device under test. The test set-up is suitable for immunity testing of electronic systems and devices according to IEC 61000-4-4, IEC 61000-4-5 and IEEE 587.

 

Standards:

IEC 61000-4-4, IEC 61000-4-5, IEEE 587 and ANSI C62.41

 

Associated Downloads

 Hilo-Test CDN 12216 Data Sheet

 

CDN 12416 Capacitive Coupling-/Decoupling Network 

CDN 12416 Capacitive Coupling-/Decoupling Network

 

Summary:

• Power supply lines

• 3* 400 V / 16 A

• Surge: 12.0 kV, 1.2/50 μs

• Surge: 6.0 kA, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

 

The Capacitive Coupling-/Decoupling Networks CDN 4416/6416/10416/12416 are used in combination with the CE-Tester or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

 

Related Standards:

IEC: IEC 61000-4-04, IEC 61000-4-05

 

 CDN 12416 Capacitive Coupling-/Decoupling Network Data Sheet
 

CDN 12432 Capacitive Coupling-/Decoupling Network 

CDN 12432 Capacitive Coupling-/Decoupling Network

 

Summary:

• Power supply lines

• 3* 400 V / 32 A

• Surge: 12.0kV, 1.2/50 μs

• Surge: 6.0A, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

 

The Capacitive Coupling-/Decoupling Networks CDN 4432/6432/10432/12432 are used in combination with the CE-Tester or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

 

Related Standards:

IEC: IEC 61000-4-04, IEC 61000-4-05

 

 CDN 12432 Capacitive Coupling-/Decoupling Network Data Sheet
 

 

CDN 12463 Capacitive Coupling-/Decoupling Network 

CDN 12463 Capacitive Coupling-/Decoupling Network

 

Summary:

• Power supply lines

• 3* 400 V / 63 A

• Surge: 12.0 kV, 1.2/50 μs

• Surge: 6.0 kA, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

 

The Capacitive Coupling-/Decoupling Networks CDN 4463/6463/10463/12463 are used in combination with the CE-Tester or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

 

 CDN 12463 Capacitive Coupling-/Decoupling Network Data Sheet
 

CDN 44100 Capacitive Coupling-/Decoupling Networks 

CDN 44100 Capacitive Coupling-/Decoupling Networks

 

Summary:

• Power supply lines

• 3* 400 V / 100 A

• Surge: 5.0 kV, 1.2/50 μs

• Surge: 2.5 kA, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

 

The Capacitive Coupling-/Decoupling Networks CDN 44100/64100/104100/124100 are used in combination with the CE-Tester or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

 

 CDN 44100 Capacitive Coupling-/Decoupling Network Data Sheet
 

Hilo-Test CDN 54100 

• Power supply lines

• 3* 400 V / 100 A

• Surge: 5.0 kV, 1.2/50 μs

• Surge: 2.5 kA, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

The capacitive Coupling-/Decoupling Networks CDN 54100/54100/104100/124100 are used in combination with the Multi-CE5 or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

Standards:

IEC 61000-4-4, IEC 61000-4-5 and IEEE 587

 

Associated Downloads

 Hilo-Test CDN 54100 Data Sheet

 

Hilo-Test CDN 54200/ CDN 74200/CDN 104200/ CDN 124200 capacitive Coupling-/Decoupling Networks 

HILO-TEST-CDN-54200-CDN-74200-CDN-104200-CDN-124200 capacitive Coupling-/Decoupling Networks

• Power supply lines

• 3* 400 V / 200 A

• Surge: 7.0 kV, 1.2/50 μs

• Surge: 3.5 kA, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

The capacitive Coupling-/Decoupling Networks CDN 54200/74200/104200/124200 are used in combination with the Multi-CE5 or the Surge generators CWG-CE7, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

Standards:

IEC 61000-4-4, IEC 61000-4-5 and IEEE 587

 

Associated Downloads

 Hilo-Test CDN 54200/ CDN 74200/CDN 104200/ CDN 124200 Data Sheet

 

Hilo-Test CDN 54250 

• Power supply lines

• 3* 400 V / 250 A

• Surge: 5.0 kV, 1.2/50 μs

• Surge: 2.5 kA, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

The capacitive Coupling-/Decoupling Networks CDN 54250/74250/104250/124250 are used in combination with the Multi-CE5 or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

Standards:

IEC 61000-4-4, IEC 61000-4-5 and IEEE 587

 

Associated Downloads

 Hilo-Test CDN 54250 Data Sheet

 

 
CDN 64100 Capacitive Coupling-/Decoupling Network 
CDN 64100 Capacitive Coupling-/Decoupling Network
 
Summary:
• Power supply lines
• 3* 400 V / 100 A
• Surge: 7.0 kV, 1.2/50 μs
• Surge: 3.5 kA, 8/20 μs
• Burst: 5.0 kV, 5/50 ns
 
The Capacitive Coupling-/Decoupling Networks CDN 44100/64100/104100/124100 are used in combination with the CE-Tester or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.
 

 

 
CDN 74100 Capacitive Coupling-/Decoupling Network 
CDN 64100 Capacitive Coupling-/Decoupling Network
 
Summary:
• Power supply lines
• 3* 400 V / 100 A
• Surge: 7.0 kV, 1.2/50 μs
• Surge: 3.5 kA, 8/20 μs
• Burst: 5.0 kV, 5/50 ns
 
The capacitive Coupling-/Decoupling Networks CDN 54100/74100/104100/124100 are used in combination with the Multi-CE5 or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.
 
Additional Info
CDN current: up to 100A
CDN voltage: up to 7kV
CDN usage: Burst (EFTG), Surge (CWG), Power supply lines
 

 

 
CDN 74250 Capacitive Coupling-/Decoupling Network 
HILO-TEST CDN 74250
 
Summary:
• Power supply lines
• 3* 400 V / 250 A
• Surge: 7.0 kV, 1.2/50 μs
• Surge: 3.5 kA, 8/20 μs
• Burst: 5.0 kV, 5/50 ns
 
The capacitive Coupling-/Decoupling Networks CDN 54250/74250/104250/124250 are used in combination with the Multi-CE5 or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.
 
Additional Info
CDN current: up to 250A
CDN voltage: up to 7kV
CDN usage: Burst (EFTG), Surge (CWG), Power supply lines
 

 

 

 

CDN 104100 Capacitive Coupling-/Decoupling Network 

CDN 104100 Capacitive Coupling-/Decoupling Network

 

Summary:

• Power supply lines

• 3* 400 V / 100 A

• Surge: 10.0 kV, 1.2/50 μs

• Surge: 5.0 kA, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

 

The Capacitive Coupling-/Decoupling Networks CDN 44100/64100/104100/124100 are used in combination with the CE-Tester or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

 

 CDN 104100 Capacitive Coupling-/Decoupling Network Data Sheet
 

 

CDN 124100 Capacitive Coupling-/Decoupling Networks 

CDN 124100 Capacitive Coupling-/Decoupling Networks

 

Summary:

• Power supply lines

• 3* 400 V / 100 A

• Surge: 12.0 kV, 1.2/50 μs

• Surge: 6.0 kA, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

 

The Capacitive Coupling-/Decoupling Networks CDN 44100/64100/104100/124100 are used in combination with the CE-Tester or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

 

 CDN 124100 Capacitive Coupling-/Decoupling Network Data Sheet
 

 

CDN 44250 Capacitive Coupling-/Decoupling Network 

 

Summary:

• Power supply lines

• 3* 400 V / 250 A

• Surge: 5.0 kV, 1.2/50 μs

• Surge: 2.5 kA, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

 

The Capacitive Coupling-/Decoupling Networks CDN 44250/64250/104250/124250 are used in combination with the CE-Tester or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

 

 CDN 44250 Capacitive Coupling-/Decoupling Network Data Sheet
 

 

CDN 64250 Capacitive Coupling-/Decoupling Network 

 

Summary:

• Power supply lines

• 3* 400 V / 250 A

• Surge: 7.0 kV, 1.2/50 μs

• Surge: 3.5 kA, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

 

The Capacitive Coupling-/Decoupling Networks CDN 44250/64250/104250/124250 are used in combination with the CE-Tester or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

 

 CDN 64250 Capacitive Coupling-/Decoupling Network Data Sheet
 

 

CDN 104250 Capacitive Coupling-/Decoupling Network 

 

Summary:

• Power supply lines

• 3* 400 V / 250 A

• Surge: 10.0 kV, 1.2/50 μs

• Surge: 5.0 kA, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

 

The Capacitive Coupling-/Decoupling Networks CDN 44250/64250/104250/124250 are used in combination with the CE-Tester or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

 

 CDN 104250 Capacitive Coupling-/Decoupling Network Data Sheet
 

 

CDN 124250 Capacitive Coupling-/Decoupling Network 

 

Summary:

• Power supply lines

• 3* 400 V / 250 A

• Surge: 12.0 kV, 1.2/50 μs

• Surge: 6.0 kA, 8/20 μs

• Burst: 5.0 kV, 5/50 ns

 

The Capacitive Coupling-/Decoupling Networks CDN 44250/64250/104250/124250 are used in combination with the CE-Tester or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

 

 CDN 124250 Capacitive Coupling-/Decoupling Network Data Sheet
 

EFTC 2012 Capacitive Coupling Clamp 

HILO-TEST EFTC 2012 Capacitive Coupling Clamp

 

Summary:

• Capacitive Coupling Clamp

• for Burst/ EFTG impulses

 

The Capacitive Coupling Clamp, EFTC 2012, provides the ability to couple Burst / EFT pulses to data lines. It can also be used for coupling to main lines where no CDN is available, for example high current lines >200 A.

 

Related Standards:

IEC: IEC 61000-4-04