EnglishPortugueseSpanish

Request a Quote / Contact Us!

    Manufacturer

    Part Number or Numbers

    HILO – Coupling Network (CDN)

    Coupling network (CDN)

    • Hilo-Test Coupling Network (CDN)

    • CDN 304 COUPLING / DECOUPLING NETWORK

    • CDN 504 sym (Symmetrical) Coupling – /Decoupling Network

    • CDN 508 sym (Symmetrical) Coupling – /Decoupling Network

    • CDN 2012 Coupling Clamp

    • CDN 2402 Capacitive Coupling-/Decoupling Network

    • CDN 2410 Capacitive Coupling-/Decoupling Network

    • CDN 2802 Capacitive Coupling-/Decoupling Network

    • CDN 4416 Capacitive Coupling-/Decoupling Networks

    • CDN 4416 B Capacitive Coupling-/Decoupling Network

    • CDN 4432 Capacitive Coupling-/Decoupling Network

    • CDN 4463 Capacitive Coupling-/Decoupling Network

    • CDN 5404 Capacitive Coupling-/Decoupling Network

    • CDN 5416 Capacitive Coupling-/Decoupling Network

    • CDN 5416 B Capacitive Coupling-/Decoupling Network

    • CDN 5432 Capacitive Coupling-/Decoupling Network

    • CDN 5463 Capacitive Coupling-/Decoupling Network

    • CDN 6416 Capacitive Coupling-/Decoupling Network

    • CDN 6432 Capacitive Coupling-/Decoupling Network

    • CDN 6463 Capacitive Coupling-/Decoupling Network

    • CDN 7416 Capacitive Coupling-/Decoupling Network

    • CDN 7432 Capacitive Coupling-/Decoupling Network

    • CDN 7463 Capacitive Coupling-/Decoupling Network

    • CDN 10216 Capacitive Coupling-/Decoupling Network

    • CDN 10416 Capacitive Coupling-/Decoupling Network

    • CDN 10432 Capacitive Coupling-/Decoupling Network

    • CDN 10463 Capacitive Coupling-/Decoupling Network

    • CDN 12216 Capacitive Coupling-/Decoupling Network

    • CDN 12416 Capacitive Coupling-/Decoupling Network

    • CDN 12432 Capacitive Coupling-/Decoupling Network

    • CDN 12463 Capacitive Coupling-/Decoupling Network

    • CDN 44100 Capacitive Coupling-/Decoupling Networks

    • CDN 44250 Capacitive Coupling-/Decoupling Network

    • CDN 54100 Capacitive Coupling-/Decoupling Network

    • CDN 54200/ CDN 74200/CDN 104200/ CDN 124200 Capacitive Coupling-/Decoupling Networks

    • CDN 54250 Capacitive Coupling-/Decoupling Network

    • CDN 64100 Capacitive Coupling-/Decoupling Network

    • CDN 64250 Capacitive Coupling-/Decoupling Network

    • CDN 74100 Capacitive Coupling-/Decoupling Network

    • CDN 74250 Capacitive Coupling-/Decoupling Network

    • CDN 104100 Capacitive Coupling-/Decoupling Network

    • CDN 104250 Capacitive Coupling-/Decoupling Network

    • CDN 124100 Capacitive Coupling-/Decoupling Networks

    • CDN 124250 Capacitive Coupling-/Decoupling Network

    • EFTC 2012 Capacitive Coupling Clamp

     

     

    Hilo-Test Coupling Network (CDN) 

    HILO-TEST offers a large number of coupling / decoupling networks (CDN) for single-phase and 3-phase power supply lines and signal data lines and telecommunication lines. In addition to offering numerous versions for current EMC Standards (IEC 61000-4-4 and IEC 61000-4-5), HILO-TEST also supplies special variants for higher voltages (e.g. 3×690 VAC or up to 1 KV DC) and high currents (up to 200 A or more), as well as for surge voltages up to 12.kV.

    The related standards are formally valid IEC 61000-4-4 and IEC 61000-4- standards.

     

    CDN 304 COUPLING / DECOUPLING NETWORK 

    Summary:

    • 100 V= / 1 A

    • max. test voltage: 5 kV, 1.2/50μs,10/700 μs

    • max. test current: 2 KA, 8/20 μs

     

    Designed for applying the impulse output voltage of the Combination Wave Generator to interconnection lines of the equipment under test, acc. to IEC 1000-4-5.

     

    Related Standards:

    IEC: IEC 61000-4-05

     

     CDN 304 Capacitive Coupling-/Decoupling Network Data Sheet
     

     

    CDN 504 sym (Symmetrical) Coupling-/Decoupling Network 

    CDN 504 sym (Symmetrical) Coupling-/Decoupling Network

    Summary:

    Rated voltage 80V=
    Surge: 5.0 kV, 1.2/50 μs
    Surge: 5.0 kV, 10/700 μs
    Coupling Impedance, selectable

     

    Designed for applying the impulse output voltage of the Combination Wave Generator to interconnection lines of the equipment under test, acc. to IEC 61000-4-5, Fig10

     

    Related Standards: IEC: IEC 61000-4-05

     

    CDN voltage: up to 5kV

    CDN usage: Surge (CWG), Data- / signal lines

     

     CDN 504 sym (Symmetrical) Coupling – /Decoupling Network Data Sheet

     

    CDN 508 sym (Symmetrical) Coupling-/Decoupling Network 

    CDN 504 sym (Symmetrical) Coupling-/Decoupling Network

    Summary:

    Rated voltage 80V=
    Surge: 5.0 kV, 1.2/50 μs
    Surge: 5.0 kV, 10/700 μs
    Coupling Impedance, selectable

     

    Designed for applying the impulse output voltage of the Combination Wave Generator to interconnection lines of the equipment under test, acc. to IEC 61000-4-5, Fig10

     

    Related Standards: IEC: IEC 61000-4-05

    CDN voltage: up to 5kV

    CDN usage: Surge (CWG), Data- / signal lines

     

     CDN 508 sym (Symmetrical) Coupling – /Decoupling Network Data Sheet

     

    CDN 2012 Coupling Clamp 

    HILO-TEST CDN 2012 Coupling Clamp

     

    Summary:

    • Coupling clamp for automotive electronics

    • Pulse injection on signal and control lines

     

    The coupling clamp CDN 2012 conforms exactly to the requirements of ISO 7637-3 and other standards, guaranteeing that tests are carried out in strict compliance with the standard. The CDN allows the fast nanosecond pulse bursts ( ISO 3a and 3b) to be injected in cable runs. Very high impedance signal lines are also to be found in vehicles on which the disruptive effects of ISO pulses 1 and 2 can also show up through Capacitive coupling.

     

    Related Standards:

    ISO: ISO 7637-3

     

     CDN 2012 Capacitive Coupling Clamp Data Sheet
     

     

    Hilo-Test CDN 2402 

    Hilo-Test CDN 2402

     

    Details:

    • Data- / signal lines
    • Surge: 2.5 kV, 1.2/50 μs
    • Burst: 2.5 kV , 5/50 ns
    • Typ. 48 V=

     

    The capacitive Coupling-/Decoupling Networks CDN 2402/2410 are used in combination with the CE-TESTER and allows superimposition of surge and burst test pulses to up to four I/O signal lines. The test set-up is suitable for surge immunity testing of electronic systems and devices according to IEC 61000-4-4, IEC 61000-4-5 and IEEE 587.

     

    Standards:

    IEC 61000-4-04, IEC 61000-4-05

     

    Associated Downloads

     Hilo-Test CDN 2402 Data Sheet

     

     

    Hilo-Test CDN 2410 

    Hilo-Test CDN 2410

     

    Details:

    • Data- / signal lines
    • Surge: 2.5 kV, 1.2/50 μs
    • Burst: 2.5 kV , 5/50 ns
    • Typ. 240V≈

     

    The capacitive Coupling-/Decoupling Networks CDN 2402/2410 are used in combination with the CE-TESTER and allows superimposition of surge and burst test pulses to up to four I/O signal lines. The test set-up is suitable for surge immunity testing of electronic systems and devices according to IEC 61000-4-4, IEC 61000-4-5 and IEEE 587.

     

    Standards:

    IEC 61000-4-04, IEC 61000-4-05

     

    Associated Downloads

     Hilo-Test CDN 2410 Data Sheet

     

     

    Hilo-Test CDN 2802 

    Hilo-Test CDN 2802

     

    Details:

    • Data- / signal lines • Surge: 2 kV, 1.2/50μs • Burst: 2 kV , 5/50 ns • 8 data lines

     

    The capacitive Coupling-/Decoupling Network CDN 2802 is used in combination with the CE-TESTER and allows superimposition of surge and burst test pulses to up to eight I/O signal lines.

     

    Standards:

    IEC 61000-4-04, IEC 61000-4-05

     

    Associated Downloads

     Hilo-Test CDN 2802 Data Sheet

     

     

    Hilo-Test CDN 4416 

    Hilo-Test CDN 4416

     

    Details:

    • Power supply lines • 3* 400 V / 16 A • Surge: 5.0 kV, 1.2/50 μs • Surge: 2.5 kA, 8/20 μs • Burst: 5.0 kV, 5/50 ns

     

    The capacitive Coupling-/Decoupling Networks CDN 4416/6416/10416/12416 are used in combination with the CE-Tester or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

     

    Standards:

    IEC 61000-4-04, IEC 61000-4-05

     

    Associated Downloads

     Hilo-Test CDN 4416 Data Sheet

     

    CDN 4416 B Capacitive Coupling-/Decoupling Network 

    CDN 4416 B capacitive Coupling-/Decoupling Network

     

    Summary:

    • Power supply lines

    • 3 * 400V / 25 A

    • Burst: 5 kV, 5/50 ns

     

    The Capacitive Coupling-/Decoupling Network CDN 4416B is used in combination with a BURST Generator EFTG 4510 or the CE-TESTER and allows superimposition burst test pulses to the 3-phase mains voltage of the device under test.

     

    Related Standards:

    IEC: IEC 61000-4-04

     

     CDN 4416 B Capacitive Coupling-/Decoupling Network Data Sheet
     

     

    Hilo-Test CDN 4432 

    HILO-TEST CDN 4432

     

    Details:

    • Power supply lines • 3* 400 V / 32 A • Surge: 5.0kV, 1.2/50 μs • Surge: 2.5A, 8/20 μs • Burst: 5.0 kV, 5/50 ns

     

    The capacitive Coupling-/Decoupling Networks CDN 4432/6432/10432/12432 are used in combination with the CE-Tester or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

     

    Standards:

    IEC 61000-4-04, IEC 61000-4-05

     

    Associated Downloads

     Hilo-Test CDN 4432 Data Sheet

     

    CDN 4463 Capacitive Coupling-/Decoupling Network 

    CDN 4463 Capacitive Coupling-/Decoupling Network

     

    Summary:

    • Power supply lines

    • 3* 400 V / 63 A

    • Surge: 5.0 kV, 1.2/50 μs

    • Surge: 2.5 kA, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

     

    The Capacitive Coupling-/Decoupling Networks CDN 4463/6463/10463/12463 are used in combination with the CE-Tester or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

     

     CDN 4463 Capacitive Coupling-/Decoupling Network Data Sheet
     

     

    HILO-TEST CDN 5404 

    Product Summary:

    • Ring wave: 4 kV, 100KHz
    • 4 data lines (2 Paar)

    The capacitive Coupling-/Decoupling Network CDN 5404 is used in combination with the IPG2554 and allows superimposition of damped oscillatory wave up to four I/O signal lines.

    For eight data lines, two devices can be used.


    Standards:
     IEC: IEC 61000-4-12, IEC 61000-4-18, IEC 60255

     

    Associated Downloads

     Hilo-Test CDN 5404 Data Sheet

     

    Hilo-Test CDN 5416 

    • Power supply lines

    • 3* 400 V / 16 A

    • Surge: 5.0 kV, 1.2/50 μs

    • Surge: 2.5 kA, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

     

    The capacitive Coupling-/Decoupling Networks CDN 5416/7416/10416/12416 are used in combination with the Multi-CE5 or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

    Standards:

    IEC 61000-4-04, IEC 61000-4-05

     

    Associated Downloads

     Hilo-Test CDN 5416 Data Sheet

     

    Hilo-Test CDN 5416 B

    • Power supply lines

    • 3* 400 V / 32 A

    • Burst: 5.0 kV, 5/50 ns

     

    The capacitive Coupling-/Decoupling Network CDN 5416B / 5432B is used in combination with a BURST Generator EFTG-CE5 or the Multi-CE5 and allows superimposition burst test pulses to the 3-phase mains voltage of the device under test.

    Standards: IEC 61000-4-04

    CDN current: up to 16A

    CDN voltage: up to 5kV

    CDN usage: Burst (EFTG), Power supply lines

     

    Associated Downloads

     Hilo-Test CDN 5416B Data Sheet

     

    Hilo-Test CDN 5432 

    • Power supply lines

    • 3* 400 V / 32 A

    • Surge: 5.0kV, 1.2/50 μs

    • Surge: 2.5A, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

     

    The capacitive Coupling-/Decoupling Networks CDN 5432/7432/10432/12432 are used in combination with the Multi-CE5 or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

    Standards:

    IEC 61000-4-04, IEC 61000-4-05

     

    Associated Downloads

     Hilo-Test CDN 5432 Data Sheet

     

     

    Hilo-Test CDN 5463 

    • Power supply lines

    • 3* 400 V / 63 A

    • Surge: 5.0 kV, 1.2/50 μs

    • Surge: 2.5 kA, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

     

    The capacitive Coupling-/Decoupling Networks CDN 5463/7463/10463/12463 are used in combination with the Multi-CE5 or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

    Standards:

    IEC 61000-4-4, IEC 61000-4-5 and IEEE 587

     

    Associated Downloads

     Hilo-Test CDN 5463 Data Sheet

     

    Hilo-Test CDN 6416

    HILO-TEST CDN 4416

     

    Details:

    • Power supply lines • 3* 400 V / 16 A • Surge: 7.0 kV, 1.2/50 μs • Surge: 3.0 kA, 8/20 μs • Burst: 5.0 kV, 5/50 ns

     

    The capacitive Coupling-/Decoupling Networks CDN 4416/6416/10416/12416 are used in combination with the CE-Tester or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

     

    Standards:

    IEC 61000-4-04, IEC 61000-4-05

     

    Associated Downloads

     Hilo-Test CDN 6416 Data Sheet

     

     

    CDN 6416 Capacitive Coupling-/Decoupling Network 

    CDN 6416 capacitive Coupling-/Decoupling Network

     

    Summary:

    • Power supply lines

    • 3* 400 V / 16 A

    • Surge: 7.0 kV, 1.2/50 μs

    • Surge: 3.0 kA, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

     

    The Capacitive Coupling-/Decoupling Networks CDN 4416/6416/10416/12416 are used in combination with the CE-Tester or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

     

    Related Standards:

    IEC: IEC 61000-4-04, IEC 61000-4-05

     

     CDN 6416 Capacitive Coupling-/Decoupling Network Data Sheet
     

     

    CDN 6432 Capacitive Coupling-/Decoupling Network 

    CDN 6432 Capacitive Coupling-/Decoupling Network

     

    Summary:

    • Power supply lines

    • 3* 400 V / 32 A

    • Surge: 7.0kV, 1.2/50 μs

    • Surge: 3.0A, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

     

    The Capacitive Coupling-/Decoupling Networks CDN 4432/6432/10432/12432 are used in combination with the CE-Tester or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

     

    Related Standards:

    IEC: IEC 61000-4-04, IEC 61000-4-05

     

     CDN 6432 Capacitive Coupling-/Decoupling Network Data Sheet
     

     

     

    CDN 6463 Capacitive Coupling-/Decoupling Network 

    CDN 6463 Capacitive Coupling-/Decoupling Network

     

    Summary:

    • Power supply lines

    • 3* 400 V / 63 A

    • Surge: 7.0 kV, 1.2/50 μs

    • Surge: 3.5 kA, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

     

    The Capacitive Coupling-/Decoupling Networks CDN 4463/6463/10463/12463 are used in combination with the CE-Tester or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

     

     CDN 6463 Capacitive Coupling-/Decoupling Network Data Sheet
     

     

     

    CDN 7416 Capacitive Coupling-/Decoupling Network 

    CDN 6463 Capacitive Coupling-/Decoupling Network

     

    Summary:

    • Power supply lines

    • 3* 400 V / 16 A

    • Surge: 7.0 kV, 1.2/50 μs

    • Surge: 3.0 kA, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

     

    The capacitive Coupling-/Decoupling Networks CDN 5416/7416/10416/12416 are used in combination with the Multi-CE5 or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

     

    Additional Info

    Standards: IEC: IEC 61000-4-04, IEC 61000-4-05

    CDN current: up to 16A

    CDN voltage: up to 7kV

    CDN usage: Burst (EFTG), Surge (CWG), Power supply lines

     

     CDN 7416 Capacitive Coupling-/Decoupling Network Data Sheet
     

     

    CDN 7432 Capacitive Coupling-/Decoupling Network 

    CDN 6463 Capacitive Coupling-/Decoupling Network

     

    Summary:

    • Power supply lines

    • 3* 400 V / 32 A

    • Surge: 7.0 kV, 1.2/50 μs

    • Surge: 3.0 kA, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

     

    The capacitive Coupling-/Decoupling Networks CDN 5432/7432/10432/12432 are used in combination with the Multi-CE5 or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

     

    Additional Info

    Standards: IEC: IEC 61000-4-04, IEC 61000-4-05

    CDN current: up to 32A

    CDN voltage: up to 7kV

    CDN usage: Burst (EFTG), Surge (CWG), Power supply lines

     

     CDN 7432 Capacitive Coupling-/Decoupling Network Data Sheet
     

    CDN 7463 Capacitive Coupling-/Decoupling Network 

    CDN 6463 Capacitive Coupling-/Decoupling Network

     

    Summary:

    • Power supply lines

    • 3* 400 V / 63 A

    • Surge: 7.0 kV, 1.2/50 μs

    • Surge: 3.5 kA, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

     

    The capacitive Coupling-/Decoupling Networks CDN 5463/7463/10463/12463 are used in combination with the Multi-CE5 or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

     

    Additional Info

    Standards:

    CDN current: up to 63A

    CDN voltage: up to 7kV

    CDN usage: Burst (EFTG), Surge (CWG), Power supply lines

     

     CDN 7463 Capacitive Coupling-/Decoupling Network Data Sheet

     

    CDN 10216 Capacitive Coupling-/Decoupling Network 

    CDN 10216 Capacitive Coupling-/Decoupling Network

    Summary:

    • Power supply lines

    • 240 V / 16 A

    • Surge: 10 kV, 1.2/50 μs

    • Surge: 5 kA, 8/20 μs

    • Burst 5 kV, 5/50 ns

     

    The Capacitive Coupling-/Decoupling Network CDN 12216 is used in combination with the Surge generators PG 7-250, PG 10-504 or PG 12-804 and allows superimposition of surge test pulses to the single-phase power supply voltage of the device under test. The test set-up is suitable for immunity testing of electronic systems and devices according to IEC 61000-4-4, IEC 61000-4-5 and IEEE 587.

     

     CDN 10216 Capacitive Coupling-/Decoupling Network Data Sheet

     

     

    Hilo-Test CDN 10416

    HILO-TEST CDN 4416

     

    Details:

    • Power supply lines • 3* 400 V / 16 A • Surge: 10.0 kV, 1.2/50 μs • Surge: 5.0 kA, 8/20 μs • Burst: 5.0 kV, 5/50 ns

     

    The capacitive Coupling-/Decoupling Networks CDN 4416/6416/10416/12416 are used in combination with the CE-Tester or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

     

    Standards:

    IEC 61000-4-04, IEC 61000-4-05

     

    Associated Downloads

     Hilo-Test CDN 10416 Data Sheet

    CDN 10432 Capacitive Coupling-/Decoupling Network 

    CDN 10432 Capacitive Coupling-/Decoupling Network

     

    Summary:

    • Power supply lines

    • 3* 400 V / 32 A

    • Surge: 10.0kV, 1.2/50 μs

    • Surge: 5.0A, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

     

    The Capacitive Coupling-/Decoupling Networks CDN 4432/6432/10432/12432 are used in combination with the CE-Tester or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

     

    Related Standards:
    IEC: IEC 61000-4-04, IEC 61000-4-05

     

     CDN 10432 Capacitive Coupling-/Decoupling Network Data Sheet
     

     

    CDN 10463 Capacitive Coupling-/Decoupling Network 

    CDN 10463 Capacitive Coupling-/Decoupling Network

     

    Summary:

    • Power supply lines

    • 3* 400 V / 63 A

    • Surge: 10.0 kV, 1.2/50 μs

    • Surge: 5.0 kA, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

     

    The Capacitive Coupling-/Decoupling Networks CDN 4463/6463/10463/12463 are used in combination with the CE-Tester or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

     

     CDN 10463 Capacitive Coupling-/Decoupling Network Data Sheet
     

     

    Hilo-Test CDN 12216

    HILO-TEST CDN 4416

     

    Details:

    • Power supply lines • 240 V / 16 A • Surge: 12 kV, 1.2/50 μs • Surge: 6 kA, 8/20 μs • Burst 5 kV, 5/50 ns

     

    The capacitive Coupling-/Decoupling Network CDN 12216 is used in combination with the Surge generators PG 7-250, PG 10-504 or PG 12-804 and allows superimposition of surge test pulses to the single-phase power supply voltage of the device under test. The test set-up is suitable for immunity testing of electronic systems and devices according to IEC 61000-4-4, IEC 61000-4-5 and IEEE 587.

     

    Standards:

    IEC 61000-4-4, IEC 61000-4-5, IEEE 587 and ANSI C62.41

     

    Associated Downloads

     Hilo-Test CDN 12216 Data Sheet

     

    CDN 12416 Capacitive Coupling-/Decoupling Network 

    CDN 12416 Capacitive Coupling-/Decoupling Network

     

    Summary:

    • Power supply lines

    • 3* 400 V / 16 A

    • Surge: 12.0 kV, 1.2/50 μs

    • Surge: 6.0 kA, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

     

    The Capacitive Coupling-/Decoupling Networks CDN 4416/6416/10416/12416 are used in combination with the CE-Tester or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

     

    Related Standards:

    IEC: IEC 61000-4-04, IEC 61000-4-05

     

     CDN 12416 Capacitive Coupling-/Decoupling Network Data Sheet
     

    CDN 12432 Capacitive Coupling-/Decoupling Network 

    CDN 12432 Capacitive Coupling-/Decoupling Network

     

    Summary:

    • Power supply lines

    • 3* 400 V / 32 A

    • Surge: 12.0kV, 1.2/50 μs

    • Surge: 6.0A, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

     

    The Capacitive Coupling-/Decoupling Networks CDN 4432/6432/10432/12432 are used in combination with the CE-Tester or the Surge generators PG 7-250, PG 10-504, PG 12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

     

    Related Standards:

    IEC: IEC 61000-4-04, IEC 61000-4-05

     

     CDN 12432 Capacitive Coupling-/Decoupling Network Data Sheet
     

     

    CDN 12463 Capacitive Coupling-/Decoupling Network 

    CDN 12463 Capacitive Coupling-/Decoupling Network

     

    Summary:

    • Power supply lines

    • 3* 400 V / 63 A

    • Surge: 12.0 kV, 1.2/50 μs

    • Surge: 6.0 kA, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

     

    The Capacitive Coupling-/Decoupling Networks CDN 4463/6463/10463/12463 are used in combination with the CE-Tester or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

     

     CDN 12463 Capacitive Coupling-/Decoupling Network Data Sheet
     

    CDN 44100 Capacitive Coupling-/Decoupling Networks 

    CDN 44100 Capacitive Coupling-/Decoupling Networks

     

    Summary:

    • Power supply lines

    • 3* 400 V / 100 A

    • Surge: 5.0 kV, 1.2/50 μs

    • Surge: 2.5 kA, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

     

    The Capacitive Coupling-/Decoupling Networks CDN 44100/64100/104100/124100 are used in combination with the CE-Tester or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

     

     CDN 44100 Capacitive Coupling-/Decoupling Network Data Sheet
     

    Hilo-Test CDN 54100 

    • Power supply lines

    • 3* 400 V / 100 A

    • Surge: 5.0 kV, 1.2/50 μs

    • Surge: 2.5 kA, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

    The capacitive Coupling-/Decoupling Networks CDN 54100/54100/104100/124100 are used in combination with the Multi-CE5 or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

    Standards:

    IEC 61000-4-4, IEC 61000-4-5 and IEEE 587

     

    Associated Downloads

     Hilo-Test CDN 54100 Data Sheet

     

    Hilo-Test CDN 54200/ CDN 74200/CDN 104200/ CDN 124200 capacitive Coupling-/Decoupling Networks 

    HILO-TEST-CDN-54200-CDN-74200-CDN-104200-CDN-124200 capacitive Coupling-/Decoupling Networks

    • Power supply lines

    • 3* 400 V / 200 A

    • Surge: 7.0 kV, 1.2/50 μs

    • Surge: 3.5 kA, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

    The capacitive Coupling-/Decoupling Networks CDN 54200/74200/104200/124200 are used in combination with the Multi-CE5 or the Surge generators CWG-CE7, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

    Standards:

    IEC 61000-4-4, IEC 61000-4-5 and IEEE 587

     

    Associated Downloads

     Hilo-Test CDN 54200/ CDN 74200/CDN 104200/ CDN 124200 Data Sheet

     

    Hilo-Test CDN 54250 

    • Power supply lines

    • 3* 400 V / 250 A

    • Surge: 5.0 kV, 1.2/50 μs

    • Surge: 2.5 kA, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

    The capacitive Coupling-/Decoupling Networks CDN 54250/74250/104250/124250 are used in combination with the Multi-CE5 or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

    Standards:

    IEC 61000-4-4, IEC 61000-4-5 and IEEE 587

     

    Associated Downloads

     Hilo-Test CDN 54250 Data Sheet

     

     
    CDN 64100 Capacitive Coupling-/Decoupling Network 
    CDN 64100 Capacitive Coupling-/Decoupling Network
     
    Summary:
    • Power supply lines
    • 3* 400 V / 100 A
    • Surge: 7.0 kV, 1.2/50 μs
    • Surge: 3.5 kA, 8/20 μs
    • Burst: 5.0 kV, 5/50 ns
     
    The Capacitive Coupling-/Decoupling Networks CDN 44100/64100/104100/124100 are used in combination with the CE-Tester or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.
     

     

     
    CDN 74100 Capacitive Coupling-/Decoupling Network 
    CDN 64100 Capacitive Coupling-/Decoupling Network
     
    Summary:
    • Power supply lines
    • 3* 400 V / 100 A
    • Surge: 7.0 kV, 1.2/50 μs
    • Surge: 3.5 kA, 8/20 μs
    • Burst: 5.0 kV, 5/50 ns
     
    The capacitive Coupling-/Decoupling Networks CDN 54100/74100/104100/124100 are used in combination with the Multi-CE5 or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.
     
    Additional Info
    CDN current: up to 100A
    CDN voltage: up to 7kV
    CDN usage: Burst (EFTG), Surge (CWG), Power supply lines
     

     

     
    CDN 74250 Capacitive Coupling-/Decoupling Network 
    HILO-TEST CDN 74250
     
    Summary:
    • Power supply lines
    • 3* 400 V / 250 A
    • Surge: 7.0 kV, 1.2/50 μs
    • Surge: 3.5 kA, 8/20 μs
    • Burst: 5.0 kV, 5/50 ns
     
    The capacitive Coupling-/Decoupling Networks CDN 54250/74250/104250/124250 are used in combination with the Multi-CE5 or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.
     
    Additional Info
    CDN current: up to 250A
    CDN voltage: up to 7kV
    CDN usage: Burst (EFTG), Surge (CWG), Power supply lines
     

     

     

     

    CDN 104100 Capacitive Coupling-/Decoupling Network 

    CDN 104100 Capacitive Coupling-/Decoupling Network

     

    Summary:

    • Power supply lines

    • 3* 400 V / 100 A

    • Surge: 10.0 kV, 1.2/50 μs

    • Surge: 5.0 kA, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

     

    The Capacitive Coupling-/Decoupling Networks CDN 44100/64100/104100/124100 are used in combination with the CE-Tester or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

     

     CDN 104100 Capacitive Coupling-/Decoupling Network Data Sheet
     

     

    CDN 124100 Capacitive Coupling-/Decoupling Networks 

    CDN 124100 Capacitive Coupling-/Decoupling Networks

     

    Summary:

    • Power supply lines

    • 3* 400 V / 100 A

    • Surge: 12.0 kV, 1.2/50 μs

    • Surge: 6.0 kA, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

     

    The Capacitive Coupling-/Decoupling Networks CDN 44100/64100/104100/124100 are used in combination with the CE-Tester or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

     

     CDN 124100 Capacitive Coupling-/Decoupling Network Data Sheet
     

     

    CDN 44250 Capacitive Coupling-/Decoupling Network 

     

    Summary:

    • Power supply lines

    • 3* 400 V / 250 A

    • Surge: 5.0 kV, 1.2/50 μs

    • Surge: 2.5 kA, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

     

    The Capacitive Coupling-/Decoupling Networks CDN 44250/64250/104250/124250 are used in combination with the CE-Tester or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

     

     CDN 44250 Capacitive Coupling-/Decoupling Network Data Sheet
     

     

    CDN 64250 Capacitive Coupling-/Decoupling Network 

     

    Summary:

    • Power supply lines

    • 3* 400 V / 250 A

    • Surge: 7.0 kV, 1.2/50 μs

    • Surge: 3.5 kA, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

     

    The Capacitive Coupling-/Decoupling Networks CDN 44250/64250/104250/124250 are used in combination with the CE-Tester or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

     

     CDN 64250 Capacitive Coupling-/Decoupling Network Data Sheet
     

     

    CDN 104250 Capacitive Coupling-/Decoupling Network 

     

    Summary:

    • Power supply lines

    • 3* 400 V / 250 A

    • Surge: 10.0 kV, 1.2/50 μs

    • Surge: 5.0 kA, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

     

    The Capacitive Coupling-/Decoupling Networks CDN 44250/64250/104250/124250 are used in combination with the CE-Tester or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

     

     CDN 104250 Capacitive Coupling-/Decoupling Network Data Sheet
     

     

    CDN 124250 Capacitive Coupling-/Decoupling Network 

     

    Summary:

    • Power supply lines

    • 3* 400 V / 250 A

    • Surge: 12.0 kV, 1.2/50 μs

    • Surge: 6.0 kA, 8/20 μs

    • Burst: 5.0 kV, 5/50 ns

     

    The Capacitive Coupling-/Decoupling Networks CDN 44250/64250/104250/124250 are used in combination with the CE-Tester or the Surge generators PG7-250, PG10-504, PG12-804 and allow superimposition of surge and burst test pulses to the 3-phase mains voltage of the device under test.

     

     CDN 124250 Capacitive Coupling-/Decoupling Network Data Sheet
     

    EFTC 2012 Capacitive Coupling Clamp 

    HILO-TEST EFTC 2012 Capacitive Coupling Clamp

     

    Summary:

    • Capacitive Coupling Clamp

    • for Burst/ EFTG impulses

     

    The Capacitive Coupling Clamp, EFTC 2012, provides the ability to couple Burst / EFT pulses to data lines. It can also be used for coupling to main lines where no CDN is available, for example high current lines >200 A.

     

    Related Standards:

    IEC: IEC 61000-4-04