HILO-TEST EMC TEST EQUIPMENT COMBINATION WAVE GENERATORS (CWG)

Data Sheet

 

Description

 

CWG Series: COMBINATION WAVE GENERATORS
Pulse voltage 1.2/50µs, Pulse current 8/20µs specified in IEC 61000-4-5, EN 61000-4-5, IEEE587, VDE0847-4-5
IPG 105 COMBINATION WAVE GENERATOR, IPG 105
Pulse voltage 1.2/50µs, – 1.0 kV, Pulse current 8/20µs, – 500 A
PG 2.5-32 COMBINATION WAVE GENERATOR, PG2.5-32
Pulse voltage 1.2/50µs, – 2,5 kV, Pulse current 8/20µs, -1.25 kA
CE-SURGE COMBINATION WAVE GENERATOR, CE-SURGE
Pulse voltage 1.2/50µs, – 5 kV, Pulse current 8/20µs, – 2.5 kA
PG 7-250 COMBINATION WAVE GENERATOR, PG 6-204
Pulse voltage 1.2/50µs, – 6 kV, Pulse current 8/20µs, – 3.0 kA
PG 10-504 COMBINATION WAVE GENERATOR, PG 10-504
Pulse voltage 1.2/50µs, – 10 kV, Pulse current 8/20µs, – 5.0 kA
PG 12-804 COMBINATION WAVE GENERATOR, PG 12-804
Pulse voltage 1.2/50µs, – 12 kV, Pulse current 8/20µs, – 6.0 kA
PG 24-2500
/ PG 24-2k5
COMBINATION WAVE GENERATOR, PG 24-2500
Pulse voltage 1.2/50µs, – 24 kV, Pulse current 8/20µs, – 12 kA
 CE-SURGE Accessories

Data Sheet

Description

 

 

Accessories for CE-SURGE

 

HI 100 Helmholtz Coil
Air coil for generation of magnetic fields according to IEC 61000-4-8 / -9
SCK 105 Surge Calibration Kit
including impulse voltage divider and current viewing resistor, used for calibration of surge generators
Coupling-/Decoupling Network for CE-SURGE 
CDN 4416
CDN 4432
CDN 4463
three-phase power supply lines 3 * 400 V, 16 A, 32 A or 63 A
max. test voltage: 5 kV, 1.2/50µs
CDN 2402 for data lines, 4 * 48 V, 2 A, max. test voltage: 2.5 kV, 1.2/50µs
CDN 2410 for data lines, 4 * 240 V, 10 A, max. test voltage: 2.5 kV, 1.2/50µs
CDN 2802 for data lines, 8 * 48 V, 10 A, max. test voltage: 5 kV, 1.2/50µs
 PG 7-250 Accessories

Data Sheet

Description

 

Accessories for PG 7-250
HI 100 Helmholtz Coil
Air coil for generation of magnetic fields according to IEC 61000-4-8 / -9
SCK 105 Surge Calibration Kit
including impulse voltage divider and current viewing resistor, used for calibration of surge generators
Coupling-/Decoupling Networks for PG 7-250 
CDN 6416
Coupling of BURST or SURGE – pulses to power supply lines 3 * 400 V, 16A,
max. Surge test voltage 7 kV, 1.2/50µs
max. Burst voltage 5 kV, 5/50ns
CDN 6432 Coupling of BURST or SURGE – pulses to power supply lines 3 * 400 V, 32A,
max. Surge test voltage 7 kV, 1.2/50µs
max. Burst voltage 5 kV, 5/50ns
CDN 6463 Coupling of BURST or SURGE – pulses to power supply lines 3 * 400 V, 63A,
max. Surge test voltage 7 kV, 1.2/50µs
max. Burst voltage 5 kV, 5/50ns
CDN 2402 Coupling of BURST or SURGE – pulses to data lines, 4 * 48 V, 2 A
CDN 2410 Coupling of BURST or SURGE – pulses to data lines, 4 * 240 V, 10 A
 PG 10-504 Accessories

Data Sheet

Description

 

 Accessories for PG 10-504 
HI 100 Helmholtz Coil
Air coil for generation of magnetic fields according to IEC 61000-4-8 / -9
SCK 105 Surge Calibration Kit
including impulse voltage divider and current viewing resistor, used for calibration of surge generators
Coupling-/Decoupling Networks for PG 10-504 
CDN 10416 Coupling of BURST or SURGE – pulses to power supply lines 3 * 400 V, 16 A
max. Surge test voltage 10 kV, 1.2/50µs
max. Burst voltage 4.5 kV, 5/50ns
CDN 10432 Coupling of BURST or SURGE – pulses to power supply lines 3 * 400 V, 32 A
max. Surge test voltage 10 kV, 1.2/50µs
max. Burst voltage 4.5 kV, 5/50ns
CDN 10463 Coupling of BURST or SURGE – pulses to power supply lines 3 * 400 V, 63 A
max. Surge test voltage 10 kV, 1.2/50µs
max. Burst voltage 4.5 kV, 5/50ns
CDN 2402 Coupling of BURST or SURGE – pulses to data lines,
4 * 48 V, 2 A bzw. 4 * 240 V, 10 A
CDN 2410 Coupling of BURST or SURGE – pulses to data lines, 4 * 240 V, 10 A
 PG 12-804 Accessories

Data Sheet

Description

 

Accessories for PG 12-804
HI 100 Helmholtz Coil
Air coil for generation of magnetic fields according to IEC 61000-4-8 / -9
SCK 105 Surge Calibration Kit
including impulse voltage divider and current viewing resistor, used for calibration of surge generators
Coupling-/Decoupling Networks for PG 12-804
CDN 12216 Coupling-/Decoupling Network
Coupling of BURST or SURGE – pulses to power supply lines 2 * 240 V, 16 A
Surge: max. test voltage: 12 kV, 1.2/50µs
coupling mode: “line to line”, “line to ground” and “common mode”
Burst:  max. test voltage: 4.5 kV, 5/50ns
according to IEC 61000-4-4 / -5, VDE 0847-4-4 / -5