    Accuracy ±1% or better, initial pulse response for all models, with a high impedance load such as 1 megOhm in parallel with 20 pF. A 50 Ohm termination will reduce the output to half. All models listed below come with a BNC connector except as noted and are to be used with a 50 Ohm coaxial cable. The current monitor case is conductive and not insulated. Adequate insulation must be provided on the conductor being monitored. To avoid electrical shock, do not mount or remove the current monitor from a live conductor.

     

     

     

    Time Domain Parameters

     

    Frequency Domain Parameters

    Model

    Shape

    Output
    (volts/amp)

    Hole
    (inches)

     

    Max. Peak
    Current
    (amps)

    Drop
    (%msec.)

    Usable
    Rise
    Time
    (nsec.)

    IT Max.
    (amp-sec.)

     

    Max
    RMS
    Curr.
    (amps)

    3dB pt.
    Low
    (Hz)

    3dB
    pt. High
    (MHz)

    I/f
    (peak
    amps
    /Hz)

    3525

    L

    0.1

    2.0

     

    5,000

    0.004

    25

    0.5

     

    100

    6.0

    15

    0.07

    4100C

    N

    1.0

    0.5

     

    500

    1.0

    15

    0.0015

     

    5

    1,500

    25

    3.6

    4100C

    N

    1.0

    0.5

     

    500

    1.0

    15

    0.0015

     

    5

    1,500

    25

    0.006

    411C

    N

    0.1

    0.5

     

    5,000

    0.015

    20

    0.15

     

    50

    25

    20

    1.5

    4160

    L

    0.01

    2.0

     

    50,000

    0.001

    200

    2.5

     

    300

    1.5

    2.0

    0.017

    4688

    L

    1.0

    2.0

     

    500

    0.4

    12

    0.005

     

    15

    600

    30

    7500

    4688

    L

    1.0

    2.0

     

    500

    0.4

    12

    0.005

     

    15

    600

    30

    3.0

    5008C

    N

    0.01

    0.5

     

    50,000

    0.005

    150

    1.0

     

    150

    7.5

    3.0

    0.6

    5101

    L

    0.5

    2.0

     

    1,000

    0.1

    20

    0.02

     

    25

    150

    18

    4.4

    5101

    L

    0.5

    2.0

     

    1,000

    0.1

    20

    0.02

     

    25

    150

    18

    0.04

    5664

    L

    0.001

    2.0

     

    200,000

    0.00025

    250

    8.0

     

    500

    0.4

    1.5

    0.25

    5949†

    K

    0.5

    4.0

     

    1,000

    0.1

    30

    0.02

     

    30

    150

    12

    12.0

    7450

    J

    0.01

    6.0

     

    50,000

    0.001

    250

    3.0

     

    400

    1.5

    1.5

    0.025

    7655

    J

    0.1

    6.0

     

    5,000

    0.007

    100

    0.4

     

    175

    10

    4.0

    1.7

    7760

    J

    1.0

    6.0

     

    500

    0.5

    50

    0.005

     

    20

    750

    7.0

    150

    7760

    J

    1.0

    6.0

     

    500

    0.5

    50

    0.005

     

    20

    750

    7.0

    0.12

    7790

    M

    1.0

    1.0

     

    500

    1.0

    15

    0.002

     

    5

    1,500

    25

    140.0

    7790

    M

    1.0

    1.0

     

    500

    1.0

    15

    0.002

     

    5

    1,500

    25

    3.5

    7795

    M

    0.1

    1.0

     

    5,000

    0.015

    25

    0.15

     

    60

    25

    15

    1.5

    7800

    M

    0.01

    1.0

     

    50,000

    0.003

    175

    0.8

     

    125

    5.0

    2.0

    3.0

    7805

    K

    1.0

    4.0

     

    500

    0.7

    25

    0.004

     

    15

    1,000

    25

    400

    7805

    K

    1.0

    4.0

     

    500

    0.7

    25

    0.004

     

    15

    1,000

    25

    0.03

    7810

    K

    0.1

    4.0

     

    5,000

    0.007

    50

    0.4

     

    150

    10

    7.0

    0.6

    7815

    K

    0.01

    4.0

     

    50,000

    0.001

    200

    3.0

     

    400

    1.5

    2.0

    0.0025

    8585C

    N

    1.0

    0.5

     

    500

    1.0

    2

    0.003

     

    5

    1,500

    200

    0.1

    8585C

    N

    1.0

    0.5

     

    500

    1.0

    2

    0.003

     

    5

    1,500

    200

    0.6

    8590C

    M

    1.0

    1.0

     

    500

    0.7

    2.5

    0.002

     

    5

    1,000

    150

    20.0

    8590C

    M

    1.0

    1.0

     

    500

    0.7

    2.5

    0.002

     

    5

    1,000

    150

    40

    8705C

    1.0

    2.0

     

    15

    4.0

    1

    .0006

     

    4

    1,500

    400

    450

    8705C

    1.0

    2.0

     

    15

    4.0

    1

    .0006

     

    4

    1,500

    400

    0.008

     

    Pearson Electronics Clamp On Current Monitors Diminsions of all models

     Accuracy ±1%, initial pulse response for all models unless otherwise stated, with a high impedance load such as 1 megOhm in parallel with 20 pF. A 50 Ohm termination will reduce the output to half.  

     Droop:   For a flat top current pulse, the output voltage decays toward zero. Initially, the decay appears linear and the slope is referred to as the droop rate. 

     Usable Rise Time: If the 10 to 90% rise time is greater than the specified usable rise time, initial overshoot will be less than 10% of the pulse amplitude.

     † =  Type N Connector