SEIBERSDORF PRODUCTS
Antennas
• PLA-R Precision Loop Antenna Receive
• PLA-SET Precision Loop Antenna Set For Site Validation
• POD 16 And POD 618 Precision Omnidirectional Dipoles Antennas
• PRD-Precision Reference Dipole
• PCD 3100 Precision Conical Dipole
• PCD 8250 – Precision Conical Dipole
Antenna Sets
• MS1 - Site VSWR Set
• MS2 - POD Antenna Set
• MS3 - MS1 Site VSWR Set without CalStan Software
• MS4 - Site VSWR Set - Automatic
• MS5 - PLA-Set for NSA validation
Antenna Site VSWR Positioners
CalStan 11.0 – RF Measurement Software
Comb Generators RefRad
Site Validation Services
• Absorber Return Loss - 200 MHz - 18 GHz
• ALSE – Absorber Lined Shielded Enclosure Measurements
• Ambient Noise Measurement
• Antenna Calibration Test Site Validation CISPR 16-1-5 30 MHz - 1 GHz
• Automotive Component Test Site Validation CISPR 25 150 KHz - 3 GHz
• Chamber FActor CENELEC R210-003 30 MHz - 1 GHz
• CTIA Ripple Test for Mobile Station 722 MHz - 2132.5 MHz
• CTIA Validation
• Field Uniformity Measurement
• Freespace VSWR Measurements
• NSA – Normalized Site Attenuation
• Shielding Effectiveness Measurement
• Site VSWR Measurement
• Table Influence
• Transmission Loss 1 Ghz - 40 Ghz
SEIBERSDORF LABORATORIES
Seibersdorf Laboratories Company
Seibersdorf Laboratories Data Sheets
Seibersdorf Laboratories Product List
RELIANT EMC PRODUCT LINES
MANUFACTURERS
Request a Quote / Contact Us!
CTIA Validation
Seibersdorf Laboratories CTIA Validation
Measurement Description
CTIA validation is done according to the “Test Plan for Mobile Station over the Air Performance”, April 2005, Revision 2.1.
Section 3: The test includes the Phi-Axis Ripple Test at 6 positions and the Theta-Axis Ripple Test at 7 positions at 2 frequencies: 836.5 MHz and 1880 MHz with 2 antenna types (dipole and loop) at each frequency. The test has to be performed with the customer’s antenna positioner. We provide calibrated antennas.
Section 4: Substitution measurement for calibrating the receive path of the test system at 2 frequencies: 836.5 MHz and 1880 MHz using our calibrated antennas.
The measurements in Section 3 and Section 4 will be done with customer’s equipment. Network analyzer, positioner, antenna adapter, cabling, test software and all other necessary test equipment has to be provided calibrated by the customer. The calibrated antennas can be provided by Seibersdorf Laboratories. If the customer aspires an accreditation it’s recommended to perform the test with customer’s antennas.
We also offer Software Verification (Appendix B and D), Uncertainty Calculation (Appendix G), Calibration of Antennas and Training & Consulting to prepare for the CTIA Audit.
Seibersdorf Laboratories Accredited re Calibration Service for EMC Test Sites Data Sheet
Seibersdorf Laboratories RF Engineering Accredited Site Validation Data Sheet