    The LISN circuit is designed and constructed to meet the performance criteria given by these standards. A LISN, or as it is also known, Artificial Mains Network (AMN), as a function of its design, performs the following functions:
    · Provides a defined, stable, impedance to the Equipment Under Test (EUT) Isolates the EUT from the power distribution network (or power source); thereby minimizing potential affects of external influences on the measurements, and facilitating accurate and repeatable results.
    · Provides a near 1:1 (<0.5 dB insertion loss) 50O coaxial measurement port (type-N) for connection to a spectrum analyzer or EMI receiver.
    Use of a Transient Limiter for impedance matching, reduction of out-of-band emissions and transient protection for your measurement instrument is highly recommended and available from Com-Power. Com-Power carries a variety of Three-Phase LISNs for CISPR 16-1-2/ANSI C63.4 of current ratings from 16 to 100 Amps. The line under test can be selected using a manual/mechanical switch or a fiber optic controlled remote interface. All Com-Power LISNs are individually calibrated in compliance with the relevant standards. Impedance, Phase, Isolation, and Insertion Loss data is supplied with each unit, along with the calibration certificate.

    • CFT-10U 10 uF Feed Through Capacitor

    • LI-1100C LISN / 150 kHz to 30 MHzCISPR 16-1-2 / ANSI C63.450 microhenry / 100 Amps

    • LI-125C LISN / 150 kHz to 30 MHzCISPR 16-1-2 / ANSI C63.450 microhenry / 25 Amps

    • LI-150C LISN / 150 kHz to 30 MHz / CISPR 16-1-2 / ANSI C63.450 microhenry / 50 Amps

    • LI-220C LISN / 9 kHz to 30 MHzCISPR 16-1-2 / ANSI C63.4250 & 50 microhenry / 20 Amps

    • LI-3100 LISN / 10 kHz to 400 MHzDO-160/ MIL-STD 461/ CISPR 25 & 16, 5 microhenry / 100 Amps
    • LI-325C LISN / 10 kHz to 400 MHzDO-160/ MIL-STD 4615 microhenry / 25 Amps

    • LI-350 LISN / 10 kHz to 400 MHzDO-160/ MIL-STD 4615 microhenry / 50 Amps

    • LI-400C LISN / 10 kHz to 10 MHz MIL-STD 461 50 microhenry / 25 Amps

    • LI-4100 / LISN 10 kHz to 10 MHzMIL-STD 46150 microhenry / 100 Amps

    • LI-550C / LISN 100 kHz to 108 MHzCISPR 25 / CISPR 16-1-25 microhenry / 50 Amps

    • LI-3P-1X / 3 Phase -LISN 150 kHz to 30 MHzCISPR 16-1-2 / ANSI C63.450

    • LI-3P-2X / 3 Phase -LISN 9 kHz to 30 MHzCISPR 16-1-2 / ANSI C63.4250 & 50

    • LI-3P-4X / 3 Phase – LISN 10 kHz to 10 MHz for MIL-STD 46

    • LIN-120C / 150 kHz to 30 MHzCISPR 16-1-2 / ANSI C63.450 microhenry / 20 Amps

    • CFT-10U 10 uF Feed Through Capacitor

    CFT-10U

    The ruggedly constructed feed-thru capacitors are highly reliable units for general use in screen rooms and other environments. It has four threaded inserts for easy installation on metal enclosure. Its M6 threaded feed-thru stud will allow power current of 100 Amps without heating or voltage loss. These have low dissipation factors, high temperature ratings, high insulation resistance, doubly rated dielectric strength, and long life. A bleeder resistor is included within the capacitors, to discharge the capacitors when applied voltage is removed.

    • Specifications:
    • · Capacitance: 10 µF ±10%
    • · Current Rating: 100 Amp at 40 degrees C
    • · Temperature: -55°C to +70°C @ rated voltage +105°C @ 50% of rated voltage
    • · Dissipation factor: Less than 1% at 25°C (measured at 120 Hz)
    • · Dielectric strength: Twice the rated VDC @ 25°C for two minutes
    • · Bleeder Insulation resistance: 1.5 MΩ + 5 % @ 25°C rated voltage.
    • · Insulation Resistance: 100MΩ prior to adding bleeder resistance
    • · Construction: Brass, Nickel Plated

    LI-1100C LISN
    150 kHz to 30 MHzCISPR 16-1-2 / ANSI C63.450 microhenry / 100 Amps

    LI-1100C

    The LI-1100 Line Impedance Stabilization Network (LISN) provides the necessary measurement platform for performing power line conducted emissions compliance testing as required by most worldwide standards for commercial products. The LI-1100 is compliant with both CISPR 16-1-2 and ANSI C63.4. The LISN provides defined stable impedance and isolates the EUT from power source influences, thereby providing accurate and repeatable results. The LI-1100 includes one pair of, separately housed, single-conductor networks, to be installed in series with each current-carrying conductor in a single- phase, dual-phase or DC power system. A second LI-1100 pair can be used to accommodate 3-phase power systems (Wye or Delta configurations). The LI-1100 is equipped with Superior Electric SUPERCON® shrouded sockets at the mains (power input) and EUT (power output) ports. The matching color-coded plugs for connection to the mains and EUT wiring are included. This LISN uses air-core inductors to prevent saturation and permeability variation. The mounting plate of the LI-1100 is left unpainted in order to facilitate connection to earth ground in its installation, which is essential due to high leakage current. Use of a Transient Limiter for impedance matching, reduction of out-of-band emissions and transient protection for your measurement instrument is highly recommended and available from Com- Power. All Com-Power LISNs are individually calibrated in compliance with the relevant requirements of CISPR 16-1-2 and ANSI C63.4. Impedance, Phase, Isolation, and Insertion Loss data is supplied with each unit, along with the calibration certificate.

     

    • Specifications:· Product Name: Line Impedance Stabilization Network (LISN)
    • · Specification: CISPR 16-1-2 / ANSI C63.4
    • · Application: Power Line conducted emission tests
    • · Frequency Range: 150KHz – 30 MHz
    • · RF Connector: 50 Ohms N-type (female)
    • · Current Rating: 100 Ampere AC, 70 Ampere DC
    • · Voltage Rating: 440 VAC (Line to Ground), 620 VDC
    • · Inductors: 50 microhenry (air-core)
    • · Mains & EUT Connections: Superior Electric SUPERCON shrouded connectors
    • · Dimensions (each network): 10 x 10 x 21 inches / 25.4 x 25.4 x 53.3 cm
    • · Weight (each network): 17 lbs / 7.7 kg
    • · Insertion Loss: <0.5 dB (150 kHz – 30 MHz)
    • · Isolation: > 40 dB (150 kHz – 30 MHz)

     

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-1100C Line Impedance Stabilization Network Data Sheet

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-1100C Line Impedance Stabilization Network Manual

     

    LI-125C LISN

    150 kHz to 30 MHzCISPR 16-1-2 / ANSI C63.450 microhenry / 25 Amps

    LI-125C

    The LI-125C Line Impedance Stabilization Network LISN provides the necessary measurement platform for performing power line conducted emissions compliance testing as required by most worldwide standards for commercial products. The LI-125C is compliant with both CISPR 16-1-2 and ANSI C63.4. The LISN provides defined stable impedance and isolates the EUT from power source influences, thereby providing accurate and repeatable results. The LI-125C includes one pair of, separately housed, single-conductor networks, to be installed in series with each current-carrying conductor in a single- phase, dual-phase or DC power system. A second LI-125C pair can be used to accommodate 3-phase power systems (Wye or Delta configurations). The LI-125C is equipped with Superior Electric SUPERCON® shrouded sockets at the mains (power input) and EUT (power output) ports. The matching color-coded plugs for connection to the mains and EUT wiring are included. This LISN uses air-core inductors to prevent saturation and permeability variation. The mounting plate of the LI-125C is left unpainted in order to facilitate connection to earth ground in its installation, which is essential due to high leakage current.

    • Specifications:
      · Product Name: Line Impedance Stabilization Network (LISN)
      · Specification: CISPR 16-1-2 / ANSI C63.4
    • · Application: Power Line conducted emission tests
    • · Frequency Range: 150KHz – 30 MHz
    • · RF Connector: 50 Ohms N-type (female)
    • · Current Rating: 25 Ampere AC, 17 Ampere DC
    • · Voltage Rating: 270 VAC (Line to Ground), 380 VDC
    • · Inductors: 50 microhenry (air-core)
    • · Mains & EUT Connections: Superior Electric SUPERCON shrouded sockets
    • · Dimensions (each network): 13 x 7 x 7 inches / 33 x 17.7 x 17.7 cm
    • · Weight (each network): 6 lbs. / 2.7 kg
    • · Insertion Loss: <0.5 dB (150 kHz – 30 MHz)
    • · Isolation: > 40 dB (150 kHz – 30 MHz)

     

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-125C LISN Data Sheet

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-125C LISN Manual

     

    LI-150C LISN

    150 kHz to 30 MHz / CISPR 16-1-2 / ANSI C63.450 microhenry / 50 Amps
    LI-150C

    This Line Impedance Stabilization Network (LISN) provides the necessary measurement platform for performing power line conducted emissions compliance testing as required by most worldwide standards for commercial products. The LI-150C is compliant with both CISPR 16-1-2 and ANSI C63.4. The LISN provides defined stable impedance and isolates the EUT from power source influences, thereby providing accurate and repeatable results. The LI-150C includes one pair of, separately housed, single-conductor networks, to be installed in series with each current-carrying conductor in a single- phase, dual-phase or DC power system. A second LI-150C pair can be used to accommodate 3-phase power systems (Wye or Delta configurations). The LI-150C is equipped with Superior Electric SUPERCON® shrouded sockets at the mains (power input) and EUT (power output) ports. The matching color-coded plugs for connection to the mains and EUT wiring are included. This LISN uses air-core inductors to prevent saturation and permeability variation. The mounting plate of the LI-150C is left unpainted in order to facilitate connection to earth ground in its installation, which is essential due to high leakage current.

    • Specifications:· Product Name: Line Impedance Stabilization Network (LISN)

      · Specification: CISPR 16-1-2 / ANSI C63.4

    • · Application: Power Line conducted emission tests
    • · Frequency Range: 150KHz – 30 MHz
    • · RF Connector: 50 Ohms N-type (female)
    • · Current Rating: 50 Ampere AC, 35 Ampere DC
    • · Voltage Rating: 270 VAC (Line to Ground), 380 VDC
    • · Inductors: 50 microhenry (air-core)
    • · Mains & EUT Connections: Superior Electric SUPERCON shrouded sockets
    • · Dimensions (each network): 15 x 10 x 10 inches / 38 x 25.4 x 25.4 cm
    • · Weight (each network): 14 lbs. / 6.3 kg
    • · Insertion Loss: <0.5 dB (150 kHz – 30 MHz)
    • · Isolation: > 40 dB (150 kHz – 30 MHz)

     

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-150C LISN Data Sheet

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-150C LISN Manual

     

    LI-220C LISN

    9 kHz to 30 MHzCISPR 16-1-2 / ANSI C63.4250 & 50 microhenry / 20 Amps
    LI-220C
    The LI-220C Line Impedance Stabilization Network (LISN) provides the necessary measurement platform for performing power line conducted emissions compliance testing as required by most worldwide standards for commercial products. The LI-220C is compliant with both CISPR 16-1-2 and ANSI C63.4. The LISN provides defined stable impedance and isolates the EUT from power source influences, thereby providing accurate and repeatable results. The LI-220C includes one pair of single-conductor networks, housed together, to be installed in series with each current-carrying conductor in a single- phase, dual-phase or DC power system. A second LI-220C pair can be used to accommodate 3-phase power systems (Wye or Delta configurations). This LISN uses air-core inductors to prevent saturation and permeability variation. The mounting plate of the LI-220C is left unpainted in order to facilitate connection to earth ground in its installation, which is essential due to high leakage current. Use of a Transient Limiter for impedance matching, reduction of out-of-band emissions and transient protection for your measurement instrument is highly recommended and available from Com- Power. All Com-Power LISNs are individually calibrated in compliance with the relevant requirements of CISPR 16-1-2 and ANSI C63.4. Impedance, Phase, Isolation, and Insertion Loss data is supplied with each unit, along with the calibration certificate.

    Specifications:

    · Product Name: Line Impedance Stabilization Network (LISN)

    • · Specification: CISPR 16-1-2 / ANSI C63.4
    • · Application: Power Line conducted emission tests
    • · Frequency Range: 9 KHz – 30 MHz
    • · RF Connector: 50 Ohms N-type (female)
    • · Current Rating: 20 Amperes AC, 14 Amperes DC
    • · Voltage Rating: 270 VAC (Line to Ground), 380 VDC
    • · Inductors: 250 & 50 microhenry (air-core)
    • · Mains & EUT Connections: IEC 60320, C-20 @Mains and Universal multi-configuration @EUT
    • · Dimensions (each network): 16.75 x 15.5 x 7.5 inches / 42.5 x 39.3 x 19 cm
    • · Weight (each network): 20.5 lbs / 9.3 kg
    • · Insertion Loss: 6 dB to 1 dB (9 kHz to 15 kHz), Decreasing linearly with log of frequency
      < 1dB (15 kHz to 30 MHz)
    • · Isolation: 20 dB to 40 dB (9kHz to 50 kHz), Increasing linearly with log of frequency
      >50 dB ( 50 kHz to 30 MHz)

     

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-220C LISN Data Sheet

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-220C LISN Manual

    LI-3100 LISN
    10 kHz to 400 MHzDO-160/ MIL-STD 461/ CISPR 25 & 16, 5 microhenry / 100 Amps

    LI-3100Line Impedance Stabilization Network (LISN) Model: LI-3100 can handle upto 100 Amps AC (70 Amps DC) current and provides the necessary measurement platform for performing power line conducted emissions compliance testing as required by DO-160 & MIL-STD 461F and CISPR 25 and CISPR 16 from 10 kHz to 400 MHz. The concept and advantages of the LISN is to: a) connect in series with, and impose a defined impedance on, the lines under test; thereby providing a distinct advantage over current or voltage probe measurements, where the impedance of the lines is largely uncontrolled; b) provide a coaxial RF measurement port for connection to the measurement equipment, to which the disturbance voltages present at the EUT port are coupled for measurement; and, c) isolate the AE (ancilliary equipment) port from the EUT and RF ports, which: 1) minimizes the influence of the AE-connected equipment/cabling on the impedance (and voltage division factor) of the network; and, 2) reduces the amplitude of any ambient signals, as well as spurious disturbance voltages generated by the AE-connected equipment, thereby minimizing their potential impact on the measurement results. Two pairs of LI-3100 can be used to accommodate 3-phase power systems (Wye or Delta configurations).

    Specifications:

    · Product Name: Line Impedance Stabilization Network (LISN)

    · Specification: RTCA DO-160 / MIL-STD 461F / CISPR 25 / CISPR 16
    · Application: Power Line conducted emission tests

    · Frequency Range: 10KHz – 400 MHz

    · RF Connector: 50 Ohms N-type (female)

    · Current Rating: 100 Amperes AC, 70 Amperes DC

    · Voltage Rating: 480 VAC (Line to Ground), 676 VDC

    · Inductors: 5 microhenry (air-core)

    · Mains & EUT Connections: Superior Electric SUPERCON Shrouded sockets

    · Dimensions (each network): 10 x 10 x 21 inches / 25.4 x 25.4 x 53.3 cm

    · Weight (each network): 13 lbs / 5.9 kg

    · Insertion Loss: <0.5 dB (100 kHz to 108 MHz)

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-3100C LISN Data Sheet

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-3100C LISN Manual

    LI-325C LISN

    10 kHz to 400 MHzDO-160/ MIL-STD 4615 microhenry / 25 Amps

    LI-325C

    LI-325C Line Impedance Stabilization Network (LISN) provides the necessary measurement platform for performing power line conducted emissions compliance testing as required by most worldwide standards for commercial products. The LI-325C is compliant with both RTCA DO-160 and MIL-STD 461. The LISN provides defined stable impedance and isolates the EUT from power source influences, thereby providing accurate and repeatable results. The LI-325C includes one pair of, separately housed, single-conductor networks, to be installed in series with each current-carrying conductor in a single- phase, dual-phase or DC power system. A second LI-325C pair can be used to accommodate 3-phase power systems (Wye or Delta configurations). The LI-325C is equipped with Superior Electric SUPERCON® shrouded sockets at the mains (power input) and EUT (power output) ports. The matching color-coded plugs for connection to the mains and EUT wiring are included. This LISN uses air-core inductors to prevent saturation and permeability variation. The mounting plate of the LI-325C is left unpainted in order to facilitate connection to earth ground in its installation, which is essential due to high leakage current. All Com-Power LISNs are individually calibrated in compliance with the relevant requirements of RTCA DO-160 and MIL-STD 461. Impedance and Insertion Loss data is supplied with each unit, along with the calibration certificate.

    Specifications:

    · Product Name: Line Impedance Stabilization Network (LISN)
    · Specification: RTCA DO-160 / MIL-STD 461D

    · Application: Power Line conducted emission tests

    · Frequency Range: 10KHz – 400 MHz RF

    · Connector: 50 Ohms N-type (female)

    · Current Rating: 25 Ampere AC, 17 Ampere DC

    · Voltage Rating: 270 VAC (Line to Ground), 380 VDC

    · Inductors: 5 microhenry (air-core)

    · Mains & EUT Connections: Superior Electric SUPERCON Shrouded sockets

    · Dimensions (each network): 13 x 7.5 x 8 inches / 33 x 19 x 20 cm · Weight (each network): 5 lbs / 2.2 kg

    · Insertion Loss: <0.3 dB (150 kHz – 30 MHz)

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-325C LISN Data Sheet

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-325C LISN Manual

    LI-350 LISN

    10 kHz to 400 MHzDO-160/ MIL-STD 4615 microhenry / 50 Amps

     LI-350

    The LI-350 LISN for DO-160 & MIL-STD 461F, can handle upto 50 Amps AC (35 Amps DC) current and provides the necessary measurement platform for performing power line conducted emissions compliance testing as required by DO-160 & MIL-STD 461F. The concept and advantages of the LISN is to: a) connect in series with, and impose a defined impedance on, the lines under test; thereby providing a distinct advantage over current or voltage probe measurements, where the impedance of the lines is largely uncontrolled; b) provide a coaxial RF measurement port for connection to the measurement equipment, to which the disturbance voltages present at the EUT port are coupled for measurement; and, c) isolate the AE (ancilliary equipment) port from the EUT and RF ports, which: 1) minimizes the influence of the AE-connected equipment/cabling on the impedance (and voltage division factor) of the network; and, 2) reduces the amplitude of any ambient signals, as well as spurious disturbance voltages generated by the AE-connected equipment, thereby minimizing their potential impact on the measurement results. Two pairs of LI-350 can be used to accommodate 3-phase power systems (Wye or Delta configurations).

    Specifications:

    · Product Name: Line Impedance Stabilization Network (LISN)
    · Specification: RTCA DO-160 / MIL-STD 461F

    · Application: Power Line conducted emission tests

    · Frequency Range: 10 kHz to 400 MHz RF

    · Connector: 50 Ohms N-type (female)

    · Current Rating: 50 Ampere AC, 35 Ampere DC

    · Voltage Rating: 480 VAC (Line to Ground), 676 VDC

    · Inductors: 5 microhenry (air-core)

    · Mains & EUT Connections: Superior Electric SUPERCON shrouded sockets

    · Dimensions (each network): 15.4 x 7 x 6.6 inches / 39.1 x 17.7 x 16.7 cm

    · Weight (each network): 6 lbs. / 2.7 kg

    · Insertion Loss: < 0.2 dB (150 kHz to 30 MHz)

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-350 LISN Data Sheet

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-350 LISN Manual

    LI-400C LISN

    10 kHz to 10 MHz MIL-STD 461 50 microhenry / 25 Amps
    LI-400C

    The LI-400C Line Impedance Stabilization Network provides the necessary measurement platform for performing power line conducted emissions compliance testing as required by most worldwide standards for commercial products. The LI-400C is compliant with MIL-STD 461. The LISN provides defined stable impedance and isolates the EUT from power source influences, thereby providing accurate and repeatable results. The LI-400C includes one pair of, separately housed, single-conductor networks, to be installed in series with each current-carrying conductor in a single- phase, dual-phase or DC power system. A second LI-400C pair can be used to accommodate 3-phase power systems (Wye or Delta configurations). The LI-400C is equipped with Superior Electric SUPERCON® shrouded sockets at the mains (power input) and EUT (power output) ports. The matching color-coded plugs for connection to the mains and EUT wiring are included. This LISN uses air-core inductors to prevent saturation and permeability variation. The mounting plate of the LI-400C is left unpainted in order to facilitate connection to earth ground in its installation, which is essential due to high leakage current. Use of a Transient Limiter for impedance matching, reduction of out-of-band emissions and transient protection for your measurement instrument is highly recommended and available from Com- Power. All Com-Power LISNs are individually calibrated in compliance with the relevant requirements of MIL-STD 461. Impedance and Insertion Loss data is supplied with each unit, along with the calibration certificate.

    Specifications:

    · Product Name: Line Impedance Stabilization Network (LISN)
    · Specification: MIL-STD 461

    · Application: Power Line conducted emission tests

    · Frequency Range: 10KHz – 10 MHz RF

    · Connector: 50 Ohms N-type (female)

    · Current Rating: 25 Amp

    · Voltage Rating: 270 VAC 50 Hz – 400 Hz (Line to Ground), 380 VDC · Inductors: 50 microhenry (air-core)

    · Mains & EUT Connections: Superior Electric SUPERCON Shrouded sockets

    · Dimensions (each network): 13 x 7 x 7 inches / 33 x 17.7 x 17.7 cm Weight (each network): 14 lbs / 6.3 kg

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-400C LISN Data Sheet

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-400C LISN Manual

    LI-4100

    LISN 10 kHz to 10 MHzMIL-STD 46150 microhenry / 100 Amps

    LI-4100

    The LI-4100 Line Impedance Stabilization Network can handle upto 100 Amps AC (70Amps DC) current and provides the necessary measurement platform for performing power line conducted emissions compliance testing as required by MIL-STD 461F. The concept and advantages of the LISN is to: a) connect in series with, and impose a defined impedance on, the lines under test; thereby providing a distinct advantage over current or voltage probe measurements, where the impedance of the lines is largely uncontrolled; b) provide a coaxial RF measurement port for connection to the measurement equipment, to which the disturbance voltages present at the EUT port are coupled for measurement; and, c) isolate the AE (ancilliary equipment) port from the EUT and RF ports, which: 1) minimizes the influence of the AE-connected equipment/cabling on the impedance (and voltage division factor) of the network; and, 2) reduces the amplitude of any ambient signals, as well as spurious disturbance voltages generated by the AE-connected equipment, thereby minimizing their potential impact on the measurement results. Two pairs of LI-4100 can be used to accommodate 3-phase power systems (Wye or Delta configurations).

    Specifications:

    · Product Name: Line Impedance Stabilization Network (LISN)

    · Specification: MIL-STD 461

    · Application: Power Line conducted emission tests

    · Frequency Range: 10 kHz to 10 MHz RF

    · Connector: 50 Ohms N-type (female)

    · Current Rating: 100 Ampere AC, 70 Ampere DC

    · Voltage Rating: 525 VAC (Line to Ground), 740 VDC

    · Inductors: 50 microhenry (air-core)

    · Mains & EUT Connections: Superior Electric SUPERCON Shrouded sockets

    · Dimensions (each network): 10 x 10 x 21 inches /25.4 x 25.4 x 53.3 cm

    · Weight (each network): 17 lbs. / 7.7 kg

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-4100 LISN Data Sheet

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-4100 LISN Manual

    LI-550C

    LISN 100 kHz to 108 MHzCISPR 25 / CISPR 16-1-25 microhenry / 50 Amps

    LI-550C

    The LI-550C Line Impedance Stabilization Network (LISN) provides the necessary measurement platform for performing power line conducted emissions compliance testing as required by most worldwide standards for commercial products. The LI-550C is compliant with both CISPR 25 and CISPR 16-1-2 . The LISN provides defined stable impedance and isolates the EUT from power source influences, thereby providing accurate and repeatable results. The LI-550C includes one pair of, separately housed, single-conductor networks, to be installed in series with each current-carrying conductor in a single- phase, dual-phase or DC power system. A second LI-550C pair can be used to accommodate 3-phase power systems (Wye or Delta configurations). The LI-550C is equipped with Superior Electric SUPERCON® shrouded sockets at the mains (power input) and EUT (power output) ports. The matching color-coded plugs for connection to the mains and EUT wiring are included. This LISN uses air-core inductors to prevent saturation and permeability variation. The mounting plate of the LI-550C is left unpainted in order to facilitate connection to earth ground in its installation, which is essential due to high leakage current.

    Specifications:

     

    · Product Name: Line Impedance Stabilization Network (LISN)
    · Specification: CISPR 25 / ANSI C16-1-2

    · Application: Power Line conducted emission tests

    · Frequency Range: 100 KHz – 108 MHz RF

    · Connector: 50 Ohms N-type (female)

    · Current Rating: 50 Ampere AC, 35 Ampere DC

    · Voltage Rating: 270 VAC (Line to Ground), 380 VDC

    · Inductors: 5 microhenry (air-core)

    · Mains & EUT Connections: Superior Electric SUPERCON shrouded sockets

    · Dimensions (each network): 13 x 7 x 7 inches / 33 x 17.7 x 17.7 cm

    · Weight (each network): 14 lbs. / 6.3 kg Insertion Loss: < 1 dB (100 kHz to 108 MHz)

    · Isolation: > 20 – 40 dB (100 kHz to 3 MHz), >40 dB (3 MHz to 30 MHz)

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-550C LISN Data Sheet

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-550C LISN Manual


    LI-3P-1X

    3 Phase -LISN 150 kHz to 30 MHzCISPR 16-1-2 / ANSI C63.450

    LI-3P-1X

    3 Phase LISN for CISPR / ANSI The LI-3P-1x series consists of four separate models of four-conductor, 50? / 50 ?H Line Impedance Stabilization Networks (LISNs). The primary differences between the four models are their respective current ratings: LI-3P-116 16 Amps (per line, continuous) LI-3P-132 32 Amps (per line, continuous) LI-3P-163 63 Amps (per line, continuous) LI-3P-1100 100 Amps (per line, continuous) The side panels for each LISN are louvered for cooling purposes. The LI-3P-163 and LI-3P-1100 also include two internal cooling fans operated by a switch on the rear panel. Three phase LISNs can be operated remotely or locally. Remote switching of the line under test (L1, L2, L3, N) is performed using the RLI-100 Remote LISN Interface, which controls the LISN via fiber optic connection. In addition to the remote method, the line under test can also be selected using the mechanical, four-position switch located on the front panel of the LISNs. Using either switching method, the lines which are not selected are internally terminated into 50 ohms, while the selected line is terminated by the 50 ohm input impedance of the measuring instrument.

    Features:

    • · Frequency Range: 150 kHz to 30 MHz
    • · Current Ratings of 16, 32, 63 and 100 Amps
    • · Fully Compliant with CISPR 16-1-2/ANSI C63.4
    • · Remote Switching of Line Under Test
    • · Four-conductor, 50Ω / 50 μH Networks appropriate for both 3Ø
      · Delta and Wye Power Configurations
    • · Three-Year Warranty

     

    (click to enlarge)

    LI-3P-1X

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-3P-1X LISN Data Sheet

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-3P-1X LISN Manual

    LI-3P-2X

    3 Phase -LISN 9 kHz to 30 MHzCISPR 16-1-2 / ANSI C63.4250 & 50

    LI-3P-2X

    The LI-3P-2x Series consists of four separate models of four-conductor, 50Ω, 50/250 μH +5Ω Line Impedance Stabilization Networks (LISNs); also known as Artificial Mains V-Networks (V-AMNs). The primary differences between the four models are their respective current ratings: LI-3P-216 16 Amps (per line, continuous) LI-3P-232 32 Amps (per line, continuous) LI-3P-263 63 Amps (per line, continuous) LI-3P-2100 100 Amps (per line, continuous)
    These LISNs provide the necessary measurement platform for performing power line conducted emissions compliance testing per most worlwide commercial EMI/EMC requirements, such as FCC (U.S.), CE (Europe), AS/NZC (Australia/New Zealand), VCCI (Japan), Industry Canada, etc.
    The LISNs perform the following functions: • provide a defned, stable power line impedance across its frequency range for the Equipment Under Test (EUT); • isolate the EUT and measurement circuit from the power source, thereby minimizing its influece on the measurements; and, • couple the disturbance voltages to the coaxial measurement port, which connects to the measuring instrument.
    The LISNs use air-core inductors to prevent saturation and permeability variation. The mounting plates are left unpainted in order to facilitate connection to earth ground in their installation, which is essential due to high leakage currents. The side panels for each LISN are louvered for cooling purposes. The LI-3P-263 and LI-3P-2100 also include two internal cooling fans operated by a switch on the rear panel.

    (click to enlarge)
    LI-3P-2X

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-3P-2X LISN Data Sheet

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-3P-2X LISN Manual


    LI-3P-4X
    3 Phase – LISN 10 kHz to 10 MHz for MIL-STD 46
    1

    LI-3P-4XThe LI-3P-4x Series consists of four separate models of four-conductor, 50Ω, 50 μH Line Impedance Stabilization Networks (LISNs). The primary differences between the four models are their respective current ratings: LI-3P-416 16 Amps (per line, continuous) LI-3P-432 32 Amps (per line, continuous) LI-3P-463 63 Amps (per line, continuous) LI-3P-4100 100 Amps (per line, continuous)
    These LISNs provide the necessary measurement platform for performing power line conducted emissions compliance testing per MIL-STD-461, CE102, conducted emissions, radio frequency potential, power leads.
    The LISNs perform the following functions: • provide a defined, stable power line impedance across its frequency range for the Equipment Under Test (EUT); • isolate the EUT and measurement circuit from the power source, thereby minimizing its influence on the measurements; and, • couple the disturbance voltages to the coaxial measurement port, which connects to the measuring instrument.
    The LISNs use air-core inductors to prevent saturation and permeability variation. The mounting plates are left unpainted in order to facilitate connection to earth ground in their installation, which is essential due to high leakage currents. The side panels for each LISN are louvered for cooling purposes. The LI-3P-463 and LI-3P-4100 also include two internal cooling fans operated by a switch on the rear panel.

    (click to enlarge)
    LI-3P-4X
    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-3P-4X LISN Data Sheet

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LI-3P-4X LISN Manual

    LIN-120C
    LISN 150 kHz to 30 MHzCISPR 16-1-2 / ANSI C63.450 microhenry / 20 Amps

    LIN-120C
    LISN 150 kHz to 30 MHz, 20 Amps The LIN-120C Line Impedance Stabilization Network (LISN) provides the necessary measurement platform for performing power line conducted emissions compliance testing as required by most worldwide standards for commercial products. The LIN-120C is compliant with both CISPR 16-1-2 and ANSI C63.4. The LISN provides defined stable impedance and isolates the EUT from power source influences, thereby providing accurate and repeatable results. The LIN-120C includes one pair of single-conductor networks, housed together, to be installed in series with each current-carrying conductor in a single-phase, dual-phase or DC power system. A second LIN-120C pair can be used to accommodate 3-phase power systems (Wye or Delta configurations).

    Specifications:

    • · Product Name: Line Impedance Stabilization Network (LISN)
    • · Specification: CISPR 16-1-2 / ANSI C63.4
    • · Application: Power Line conducted emission tests
    • · Frequency Range: 150KHz – 30 MHz
    • · RF Connector: 50 Ohms N-type (female)
    • · Current Rating: 20 Amperes AC, 14 Amperes DC
    • · Voltage Rating: 270 VAC (Line to Ground), 380 VDC
    • · Inductors: 50 microhenry (air-core)
    • · Mains & EUT Connections: IEC 60320, C-20 @Mains and Universal multi-configuration @ EUT
    • · Dimensions (each network): 14 x 13 x 6.5 inches / 35.5 x 33 x 16.5 cm
    • · Weight (each network): 10.5 lbs / 4.76 kg
    • · Insertion Loss: <0.5 dB (150 kHz – 30 MHz)
    • · Isolation: > 40 dB (150 kHz – 30 MHz)

    Aaronia_AG_EMC_Analyzer_Bundle-1_Data_Sheet Com-Power LIN-120C LISN Data Sheet